A compactly integrated flow cell with a chemiluminescent FIA system for determining lactate concentration in serum

被引:70
作者
Nakamura, H
Murakami, Y
Yokoyama, K
Tamiya, E
Karube, I
机构
[1] Univ Tokyo, Adv Sci & Technol Res Ctr, Meguro Ku, Tokyo 1538904, Japan
[2] Seiko Instruments Inc, Corp R&D, Chiba 2702222, Japan
[3] Saitama Inst Technol, Dept Environm Engn, Okabe, Saitama 3690203, Japan
关键词
D O I
10.1021/ac000855u
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We have fabricated an integrated now cell as a total microanalysis system (mu TAS). This flow cell (size, 15 x 20 mm; total inner volume, 12.2 muL) was designed for a rational analyzing system of lactate determination for serum. This cell was made by micromachining techniques and consisted of two hollows of a lactate oxidase (LOD) reactor and a mixing cell, a spiral groove, and three penetrated holes. To form the reactor and capillary, these patterns, etched on a silicon wafer, were attached to a glass plate by the anodic bonding method. A photodiode was put under part of the spiral capillary. The compactly accumulated devices were integrated into a now injection analysis (FIA) system. In the now cell, lactate was catalyzed to pyruvate and hydrogen peroxide at the LOD reactor; subsequently, hydrogen peroxide reacted with the luminol-ferricyanic reagent at the mixing cell. The resulting chemiluminescent light was detected by the photodiode. Using the miniaturized now cell, the sample volume for one measurement was greatly reduced to 0.2 muL. The response to lactate was obtained within 30s and was linear between 0.5 and 5.0 mM (4.5 and 45 mg/dL) lactate with excellent correlative variances of 3.2% (average of three measurements at 5.0 mM), For practical application, the lactate concentration in control human serum was determined using this system. The results showed a good correlation coefficient (r = 0.979) with the results obtained by the spectrophotometric reference method. No difference in sera (normal or pathological) was found. Consequently, this integrated now cell shows potential as a clinical device for lactate determination in serum. In this article, the effect of the design on the chemiluminescent FIA system is also described.
引用
收藏
页码:373 / 378
页数:6
相关论文
共 27 条
[1]   A microfabricated flow-through cell with parallel-opposed electrodes for recycling amperometric detection [J].
Brooks, SA ;
Kennedy, RT .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 1997, 436 (1-2) :27-34
[2]  
GIRERER RJ, 1998, BIOSENS BIOELECTRON, V13, P1271
[3]  
GUTMANN I, 1974, METHODS ENZYMATIC AN
[4]   Novel measurement of hypoxanthine in fish using direct measurement probe and chemiluminescence flow injection analysis [J].
Hayashi, K ;
Okugawa, T ;
Kozuka, Y ;
Sasaki, S ;
Ikebukuro, K ;
Karube, I .
JOURNAL OF FOOD SCIENCE, 1996, 61 (04) :736-&
[5]   A novel biosensor system for cyanide based on a chemiluminescence reaction [J].
Ikebukuro, K ;
Shimomura, M ;
Onuma, N ;
Watanabe, A ;
Nomura, Y ;
Nakanishi, K ;
Arikawa, Y ;
Karube, I .
ANALYTICA CHIMICA ACTA, 1996, 329 (1-2) :111-116
[6]  
Krawczyk TKV, 1996, BIOSENS BIOELECTRON, V11, P1155, DOI 10.1016/0956-5663(96)82338-4
[7]   Luminol electrochemiluminescence-based fibre optic biosensors for flow injection analysis of glucose and lactate in natural samples [J].
Marquette, CA ;
Blum, LJ .
ANALYTICA CHIMICA ACTA, 1999, 381 (01) :1-10
[8]   INTEGRATION OF ENZYME-IMMOBILIZED COLUMN WITH ELECTROCHEMICAL FLOW CELL USING MICROMACHINING TECHNIQUES FOR A GLUCOSE DETECTION SYSTEM [J].
MURAKAMI, Y ;
TAKEUCHI, T ;
YOKOYAMA, K ;
TAMIYA, E ;
KARUBE, I ;
SUDA, M .
ANALYTICAL CHEMISTRY, 1993, 65 (20) :2731-2735
[9]   MICROMACHINED ELECTROCHEMICAL FLOW CELL FOR BIOSENSING [J].
MURAKAMI, Y ;
UCHIDA, T ;
TAKEUCHI, T ;
TAMIYA, E ;
KARUBE, I ;
SUDA, M .
ELECTROANALYSIS, 1994, 6 (09) :735-739
[10]   MICROMACHINED ENZYME REACTOR FOR FIA SYSTEM [J].
MURAKAMI, Y ;
SUDA, M ;
YOKOYAMA, K ;
TAKEUCHI, T ;
TAMIYA, E ;
KARUBE, I .
MICROCHEMICAL JOURNAL, 1994, 49 (2-3) :319-325