Adaptive temporal resolution optimization in helical cardiac cone beam CT reconstruction

被引:93
作者
Manzke, R [1 ]
Grass, M
Nielsen, T
Shechter, G
Hawkes, D
机构
[1] Philips Res Labs, Sector Tech Syst, Hamburg, Germany
[2] KCL, Imaging Sci Grp, London, England
[3] Philips Med Syst, Haifa, Israel
关键词
cardiac cone beam computed tomography; temporal resolution; optimization of approximate 3D reconstruction;
D O I
10.1118/1.1624756
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Cone beam computed tomography scanners in combination with heart rate adaptive reconstruction schemes have the potential to enable cardiac volumetric computed tomography (CT) imaging for a larger number of patients and applications. In this publication, an adaptive scheme for the automatic and patient-specific reconstruction optimization is introduced to improve the temporal resolution and image quality. The optimization method permits the automatic determination of the required amount of gated helical cone beam projection data for the reconstruction volume. It furthermore allows one to optimize subvolume reconstruction yielding an increased temporal resolution. In addition, methods for the assessment of the temporal resolution are given which enable a quantitative documentation of the reconstruction improvements. Results are presented for patient data sets acquired in low pitch helical mode using a 16-slice cone beam CT system with parallel ECG recording. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:3072 / 3080
页数:9
相关论文
共 25 条
[1]  
Cline H, 2000, Circulation, V102, P1589
[2]  
DANIELSSON PE, 1997, P 4 INT C FULL 3D RE
[3]   Rebinning-based algorithms for helical cone-beam CT [J].
Defrise, M ;
Noo, F ;
Kudo, H .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (11) :2911-2937
[4]   Heart rate adaptive optimization of spatial and temporal resolution for electrocardiogram-gated multislice spiral CT of the heart [J].
Flohr, T ;
Ohnesorge, B .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2001, 25 (06) :907-923
[5]   Helical cardiac cone beam reconstruction using retrospective ECG gating [J].
Grass, M ;
Manzke, R ;
Nielsen, T ;
Koken, P ;
Proksa, R ;
Natanzon, M ;
Shechter, G .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (18) :3069-3084
[6]  
HOFFMANN M, UNPUB RADIOLOGY
[7]   Multislice helical CT: Image temporal resolution [J].
Hu, H ;
Pan, TS ;
Shen, Y .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2000, 19 (05) :384-390
[8]  
Kachelriess M, 2000, IEEE T MED IMAGING, V19, P888, DOI 10.1109/42.887837
[9]   ECG-correlated image reconstruction from subsecond multi-slice spiral CT scans of the heart [J].
Kachelriess, M ;
Ulzheimer, S ;
Kalender, WA .
MEDICAL PHYSICS, 2000, 27 (08) :1881-1902
[10]   Electrocardiogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart [J].
Kachelriess, M ;
Kalender, WA .
MEDICAL PHYSICS, 1998, 25 (12) :2417-2431