Evaluation of radiation exposure with singleslice- and a multislice-spiral CT system (a phantom study)

被引:26
作者
Giacomuzzi, SM
Torbica, P
Rieger, M
Lottersberger, C
Peer, S
Peer, R
Perkmann, R
Buchberger, W
Bale, R
Mallouhi, A
Jaschke, W
机构
[1] Univ Klin Radiodiagnost, Innsbruck, Austria
[2] Inst Med Phys, Innsbruck, Austria
[3] Univ Klin Gefasschirurg, Innsbruck, Austria
来源
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN | 2001年 / 173卷 / 07期
关键词
multislice spiral-CT; dose; Alderson-Rando-Phantom; single slice spiral-CT; computed tomography;
D O I
10.1055/s-2001-15831
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of study was to compare patient dose applying singleslice- and multislice-spiral CT. Methods: The examinations were performed with a singleslice-spiral CT (Highspeed Advantage, GE Medical Systems; Milwaukee, USA) and with a multislice CT systems (LightSpeed QX/i GE Medical Systems; Milwaukee, USA). For the determination of the radiation exposure (absorbed dose) a selection of most executed protocols (thorax-helical, abdomen-helical, petrous bone-axial, head-axial) were simulated using an Alderson Rando Phantom. The dose was determined by means of lithiumfluorid-thermoluminiscence dosimeters (TLD-GR 200). Results: For thorax and abdomen protocols higher energy dose values could be found using a multislice CT. On the average the energy dose values were increased by 2.6 on an average in relation to single slice spiral CT. The energy dose values of the multisclice CT using head protocols could be reduced by 30 % in relation to single slice spiral CT due to suitable parameter selections. The energy dose applying a petrous bone protocol resulted in an average increase by a factor 1.5 using a multislice CT. Conclusion: Using the new multislice CT technique protocol strategies must be optimized regarding the patient doses. Users can operate critically in the sense of the radiation protection only if they are aware of the occurring dose amounts to the patient.
引用
收藏
页码:643 / 649
页数:7
相关论文
共 33 条
[1]  
Angerstein W, 1995, Rontgenpraxis, V48, P261
[2]  
Becker CR, 1998, RADIOLOGE, V38, P726, DOI 10.1007/s001170050416
[3]  
BERNHARDT JH, 1995, STRAHLENEXPOSITION M, V30, P179
[4]   Dosage measurement in spiral CT examinations in the head and neck region [J].
Cohnen, M ;
Cohnen, B ;
Ewen, K ;
Teubert, G ;
Modder, U .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1998, 168 (05) :474-479
[5]   Multislice spiral CT of the paranasal sinuses. [J].
Dammann, F ;
Bode, A ;
Heuschmid, M ;
Kopp, A ;
Georg, C ;
Pereira, PL ;
Claussen, CD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (08) :701-706
[6]   Radiation exposure with spiral CT of the paranasal sinuses [J].
Dammann, F ;
Momino-Traserra, E ;
Remy, C ;
Pereira, PL ;
Baumann, I ;
Koitschev, A ;
Claussen, CD .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2000, 172 (03) :232-237
[7]   The Smart-Scan-Technique in spiral computed tomography. A new method for dose reduction. [J].
Giacomuzzi, SM ;
Erckert, B ;
Schopf, T ;
Freund, MC ;
Springer, P ;
Dessl, A ;
Jaschke, W .
ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 1996, 165 (01) :10-16
[8]   Dose reduction in CT by anatomically adapted tube current modulation. 1. Simulation studies [J].
Gies, M ;
Kalender, WA ;
Wolf, H ;
Suess, C ;
Madsen, MT .
MEDICAL PHYSICS, 1999, 26 (11) :2235-2247
[9]  
JANECZEK J, 1995, RADIAT PROT DOSIM, V60, P231
[10]   ANATOMICALLY ADAPTED VARIATION OF THE TUBE CURRENT IN CT - STUDIES ON REDUCTION OF RADIATION-DOSE AND IMAGE QUALITY [J].
KOPKA, L ;
FUNKE, M ;
BREITER, N ;
HERMANN, KP ;
VOSSHENRICH, R ;
GRABBE, E .
FORTSCHRITTE AUF DEM GEBIETE DER RONTGENSTRAHLEN UND DER NEUEN BILDGEBENDEN VERFAHREN, 1995, 163 (05) :383-387