Computer-Assisted Techniques Versus Conventional Guides for Component Alignment in Total Knee Arthroplasty

被引:100
作者
Blakeney, William G. [1 ]
Khan, Riaz J. K.
Wall, Simon J.
机构
[1] Sir Charles Gairdner Hosp, Dept Orthoped Surg, Nedlands, WA 6009, Australia
关键词
TIBIAL ALIGNMENT; PROSPECTIVE TRIAL; REPLACEMENT; INTRAMEDULLARY; EXTRAMEDULLARY; SYSTEMS; MECHANISMS; NAVIGATION; PLACEMENT; FAILURE;
D O I
10.2106/JBJS.I.01321
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
100224 [整形外科学];
摘要
Background: Optimal alignment of the prosthesis in total knee arthroplasty results in improved patient outcomes. The goal of this study was to determine the most accurate technique for component alignment in total knee arthroplasty by comparing computer-assisted surgery with two conventional techniques involving use of an intramedullary guide for the femur and either an intramedullary or an extramedullary guide for the tibia. Methods: One hundred and seven patients were randomized prior to surgery to one of three arms: computer-assisted surgery for both the femur and the tibia (the computer-assisted surgery group), intramedullary guides for both the femur and the tibia (the intramedullary guide group), and an intramedullary guide for the femur and an extramedullary guide for the tibia (the extramedullary guide group). Measurements of alignment on hip-to-ankle radiographs and computed tomography (CT) scans made three months after surgery were evaluated. The operative times and complications were compared among the three groups. Results: The coronal tibiofemoral angle demonstrated, on average, less malalignment in the computer-assisted surgery group (1.91 degrees) than in the extramedullary (3.22 degrees) and intramedullary (2.59 degrees) groups (p = 0.007). The coronal tibiofemoral angle was >3 degrees of varus or valgus deviation in 19% (seven) of the thirty-six patients treated with computer-assisted surgery compared with 38% (thirteen) of the thirty-four in the extramedullary guide group and 36% (thirteen) of the thirty-six in the intramedullary guide group (p = 0.022). The increase in accuracy with computer-assisted surgery came at a cost of increased operative time. The operative time for the computer-assisted surgery group averaged 107 minutes compared with eighty-three and eighty minutes, respectively, for the surgery with the extramedullary and intramedullary guides (p < 0.0001). There was no significant difference in any of the outcomes between the intramedullary and extramedullary guide groups. Conclusions: This study provides evidence that the implant alignment with computer-assisted total knee arthroplasty, as measured with radiography and computed tomography, is significantly improved compared with that associated with conventional surgery with intramedullary or extramedullary guides. This finding adds to the body of evidence showing an improved radiographic outcome with computer-assisted surgery compared with that following conventional total knee arthroplasty.
引用
收藏
页码:1377 / 1384
页数:8
相关论文
共 30 条
[1]
BARGREN JH, 1983, CLIN ORTHOP RELAT R, P178
[2]
Navigated total knee replacement - A meta-analysis [J].
Bauwens, Kai ;
Matthes, Gerrit ;
Wich, Michael ;
Gebhard, Florian ;
Hanson, Beate ;
Ekkernkamp, Axel ;
Stengel, Dirk .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (02) :261-269
[3]
Tibial component failure mechanisms in total knee arthroplasty [J].
Berend, ME ;
Ritter, MA ;
Meding, JB ;
Faris, PM ;
Keating, EM ;
Redelman, R ;
Faris, GW ;
Davis, KE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (428) :26-34
[4]
Malrotation causing patellofemoral complications after total knee arthroplasty [J].
Berger, RA ;
Crossett, LS ;
Jacobs, JJ ;
Rubash, HE .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1998, (356) :144-153
[5]
BRYS DA, 1991, CLIN ORTHOP RELAT R, P175
[6]
Computer-assisted total knee replacement - A controlled cadaver study using a multi-parameter quantitative CT assessment of alignment (the Perth CT protocol) [J].
Chauhan, SK ;
Clark, GW ;
Lloyd, S ;
Scott, RG ;
Breidahl, W ;
Sikorski, JM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (06) :818-823
[7]
Computer-assisted knee arthroplasty versus a conventional jig-based technique -: A randomised, prospective trial [J].
Chauhan, SK ;
Scott, RG ;
Breidahl, W ;
Beaver, RJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2004, 86B (03) :372-377
[8]
Randomized control trial comparing radiographic total knee arthroplasty implant placement using computer navigation versus conventional technique [J].
Chin, PL ;
Yang, KY ;
Yeo, SJ ;
Lo, NN .
JOURNAL OF ARTHROPLASTY, 2005, 20 (05) :618-626
[9]
Does Accurate Anatomical Alignment Result in Better Function and Quality of Life? Comparing Conventional and Computer-Assisted Total Knee Arthroplasty [J].
Choong, Peter F. ;
Dowsey, Michelle M. ;
Stoney, James D. .
JOURNAL OF ARTHROPLASTY, 2009, 24 (04) :560-569
[10]
Leg axis after computer-navigated total knee arthroplasty - A prospective randomized trial comparing computer-navigated and manual implantation [J].
Decking, R ;
Markmann, Y ;
Fuchs, J ;
Puhl, W ;
Scharf, HP .
JOURNAL OF ARTHROPLASTY, 2005, 20 (03) :282-288