Simultaneous and rapid analysis of cyclosporin A and creatinine in finger prick blood samples using liquid chromatography tandem mass spectrometry and its application in C2 monitoring

被引:43
作者
Keevil, BG
Tierney, DP
Cooper, DP
Morris, MR
Machaal, A
Yonan, N
机构
[1] Wythenshawe Hosp, Dept Clin Biochem, Manchester M23 9LT, Lancs, England
[2] Wythenshawe Hosp, Transplant Unit, Manchester M23 9LT, Lancs, England
[3] Micromass UK Ltd, Clin Applicat Grp, Manchester, Lancs, England
关键词
cyclosporin; tandem mass spectrometry;
D O I
10.1097/00007691-200212000-00013
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
A simple and rapid liquid chromatography tandem mass spectrometry (LC-MS/MS) method for the simultaneous analysis of cyclosporin A (CsA) and creatinine using capillary blood has been developed. Venous and capillary blood samples were taken predose and at C2 from 65 heart and lung transplant recipients (65 x 4 samples). For comparisons, serum creatinine and blood CsA concentrations were measured by the Jaffe and EMIT methods, respectively, using an Olympus AU600 analyzer. For the LC-MS/MS assay, samples were prepared in a 96 x 700-muL well block by adding 10 muL of blood (or serum) to 40 muL of 0.1 mol/L zinc sulphate solution containing deuterated creatinine internal standard. Proteins were precipitated by adding 100 muL acetonitrile containing ascomycin internal standard. After vigorous mixing and centrifugation, 5 [muL of the supernatant was injected into the LC-MS/MS system. A Waters 2795 high-performance liquid chromatography (HPLC system was used to elute a C 18 cartridge (3 mm x 4 mm) at 0.6 mL/min with a step gradient of 50-100% methanol containing 2 mmol/L ammonium acetate and 0.1% (v/v) formic acid. The column was maintained at 55degreesC, and the retention times were creatinine, 0.4 minutes; ascomycin, 0.98 minutes; and CsA, 1.2 minutes. Cycle time was 2.5 minutes, injection to injection. The analytes were monitored using a Quattro microtandem mass spectrometer operated in multiple reaction monitoring mode using the following transitions: creatinine, m/z, 1 114>44; d3-creatinine (IS), m/z, 117>47; ascomycin (IS), m/z 809>756; and CsA, m/z, 1,220>1,203. Assay characteristics were CsA intraassay CV, 3.6-3.0% (33-1,500 mug/L); CsA interassay CV, 6.7-2.5% (10-5,000 mug/L); LC-MS/MS capillary [CsA] = 0.99 x LC-MS/MS venous [CsA] - 4.2, R = 0.98; and LC-MS/MS venous [CsA] = 0.93 x EMIT venous [CsA] + 2.9, R = 0.98. Creatinine intraassay CV, 6.6-2.5% (20-720 mumol/L); interassay CV, 5.7-3.3% (80-590 mumol/L); LC-MS/MS capillary [creatinine] = 0.99 Jaffe plasma [creatinine] -42.6, R = 0.87. Total time for the preparation and analysis of 30 samples was approximately 2 hours. This assay will provide a flexible, robust, and cost-effective solution for the monitoring of CsA and creatinine in transplant recipients with potential applications in pediatric medicine and pharmacokinetic studies, in which frequent sampling is required.
引用
收藏
页码:757 / 767
页数:11
相关论文
共 26 条
[11]  
Keevil BG, 2002, CLIN CHEM, V48, P69
[12]   Simultaneous on-line extraction and analysis of sirolimus (rapamycin) and ciclosporin in blood by liquid chromatography electrospray mass spectrometry [J].
Kirchner, GI ;
Vidal, C ;
Jacobsen, W ;
Franzke, A ;
Hallensleben, K ;
Christians, U ;
Sewing, KF .
JOURNAL OF CHROMATOGRAPHY B, 1999, 721 (02) :285-294
[13]  
LEAKER B, 1994, BRIT J HOSP MED, V52, P529
[14]   C2 monitoring strategy for optimising cyclosporin immunosuppression from the neoral®1 formulation [J].
Levy, GA .
BIODRUGS, 2001, 15 (05) :279-290
[15]   INFLUENCE OF CYCLOSPORINE PHARMACOKINETICS, TROUGH CONCENTRATIONS, AND AUC MONITORING ON OUTCOME AFTER KIDNEY-TRANSPLANTATION [J].
LINDHOLM, A ;
KAHAN, BD .
CLINICAL PHARMACOLOGY & THERAPEUTICS, 1993, 54 (02) :205-218
[16]   Quantitation of cyclosporin A in whole blood by liquid chromatography/stable isotope dilution electrospray ionization mass spectrometry [J].
Magni, F ;
Pereira, S ;
Leoni, M ;
Grisenti, G ;
Kienle, MG .
JOURNAL OF MASS SPECTROMETRY, 2001, 36 (06) :670-676
[17]   Neoral monitoring by simplified sparse sampling area under the concentration-time curve - Its relationship to acute rejection and cyclosporine nephrotoxicity early after kidney transplantation [J].
Mahalati, K ;
Belitsky, P ;
Sketris, I ;
West, K ;
Panek, R .
TRANSPLANTATION, 1999, 68 (01) :55-62
[18]   Accuracy of cyclosporin measurements made in capillary blood samples obtained by skin puncture [J].
Merton, G ;
Jones, K ;
Lee, M ;
Johnston, A ;
Holt, DW .
THERAPEUTIC DRUG MONITORING, 2000, 22 (05) :594-598
[19]  
MEYER MM, 1993, J AM SOC NEPHROL, V4, P1306
[20]   LAKE-LOUISE-CONSENSUS-CONFERENCE ON CYCLOSPORINE ONITORING IN ORGAN-TRANSPLANTATION - REPORT OF THE CONSENSUS PANEL [J].
OELLERICH, M ;
ARMSTRONG, VW ;
KAHAN, B ;
SHAW, L ;
HOLT, DW ;
YATSCOFF, R ;
LINDHOLM, A ;
HALLORAN, P ;
GALLICANO, K ;
WONIGEIT, K ;
SCHUTZ, E ;
SCHRAN, H ;
ANNESLEY, T .
THERAPEUTIC DRUG MONITORING, 1995, 17 (06) :642-654