Compatibility of interventional x-ray and magnetic resonance imaging: Feasibility of a closed bore XMR (CBXMR) system

被引:18
作者
Brzozowski, Lukasz
Ganguly, Arundhuti
Pop, Mihaela
Wen, Zhifei
Bennett, Robert
Fahrig, Rebecca
Rowlands, J. A.
机构
[1] Univ Toronto, Sunnybrook Hlth Sci Ctr, Dept Med Biophys, Toronto, ON M4N 3M5, Canada
[2] Stanford Univ, Dept Radiol, Lucas MRS Ctr, Stanford, CA 94305 USA
关键词
MRI; magnetic resonance imaging; hybrid MRI; magnetic field; x-ray tube; GUIDED PLACEMENT; MR; FLUOROSCOPY; FIELD;
D O I
10.1118/1.2219328
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A next-generation interventional guidance system is proposed that will enable intraprocedural access to both x-ray and magnetic resonance imaging (MRI) modalities. This closed bore XMR (CBXMR) system is comprised of a conventional radiographic rotating anode x-ray tube and a direct conversion flat panel detector on a rotating gantry positioned adjacent to the bore of a 1.5 T MRI. To assess the feasibility of such a system, we have investigated the degree of compatibility between the x-ray components and the MRI. For \(B) over right arrow (ext)\ < < 200 G the effect on the radiographic tube motor was negligible regardless of the orientation of B, with respect to the motor axis of rotation-the frequency of anode rotation remained within 6% of the 3400 rpm frequency measured at 0 G. For \(B) over right arrow (ext)\> 2400 G the anode slowed down to below 2400 rpm at all orientations. At intermediate the frequency of rotation varied between 2400 and 3200 rpm, showing a strong dependence on orientation, with (B) over right arrow (ext) perpendicular to the tube axis having a much stronger effect on the rotation frequency than (B) over right arrow (ext) parallel to the tube axis. In contrast to the effect of (B) over right arrow (ext) on the induction motor, parallel (B) over right arrow (ext) had a stronger detrimental effect on the cathode-anode electron beam, whose path was at 16 degrees to the tube axis, than the perpendicular (B) over right arrow (ext). Parallel (B) over right arrow (ext) of several hundred Gauss had a defocusing effect on the x-ray focal spot. (B) over right arrow (ext) perpendicular to the electron beam shifted the beam without significant defocusing. We have determined that the direct conversion flat panel detector (FPD) technology is not intrinsically sensitive to (B) over right arrow (ext), and that the modifications required to make the proposed FPDs MRI compatible are minimal. The homogeneity of the MRI signal in the normal field of view was not significantly degraded by the presence of these x-ray components in the vicinity of the MRI bore entrance. (C) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:3033 / 3045
页数:13
相关论文
共 22 条
[1]   Magnetic resonance-guided placement of atrial septal closure device in animal model of patent foramen ovale [J].
Buecker, A ;
Spuentrup, E ;
Grabitz, R ;
Freudenthal, F ;
Muehler, EG ;
Schaeffter, T ;
van Vaals, JJ ;
Günther, RW .
CIRCULATION, 2002, 106 (04) :511-515
[2]  
Buecker A, 2000, J MAGN RESON IMAGING, V12, P616, DOI 10.1002/1522-2586(200010)12:4<616::AID-JMRI15>3.0.CO
[3]  
2-F
[4]   Magnetic resonance fluoroscopy allows targeted delivery of mesenchymal stem cells to infarct borders in swine [J].
Dick, AJ ;
Guttman, MA ;
Raman, VK ;
Peters, DC ;
Pessanha, BSS ;
Hill, JM ;
Smith, S ;
Scott, G ;
McVeigh, ER ;
Lederman, RJ .
CIRCULATION, 2003, 108 (23) :2899-2904
[5]   Invasive human magnetic resonance imaging: Feasibility during revascularization in a combined XMR suite [J].
Dick, AJ ;
Raman, VK ;
Raval, AN ;
Guttman, MA ;
Thompson, RB ;
Ozturk, C ;
Peters, DC ;
Stine, AM ;
Wright, VJ ;
Schenke, WH ;
Lederman, RJ .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2005, 64 (03) :265-274
[6]   Performance of a static-anode/flat-panel x-ray fluoroscopy system in a diagnostic strength magnetic field: A truly hybrid x-ray/MR imaging system [J].
Fahrig, R ;
Wen, Z ;
Ganguly, A ;
DeCrescenzo, G ;
Rowlands, JA ;
Stevens, GM ;
Saunders, RF ;
Pelc, NJ .
MEDICAL PHYSICS, 2005, 32 (06) :1775-1784
[7]   First use of a truly-hybrid X-ray/MR imaging system for guidance of brain biopsy [J].
Fahrig, R ;
Heit, G ;
Wen, Z ;
Daniel, BL ;
Butts, K ;
Pelc, NJ .
ACTA NEUROCHIRURGICA, 2003, 145 (11) :995-997
[8]   Truly hybrid interventional MR/x-ray system: Investigation of in vivo applications [J].
Fahrig, R ;
Butts, K ;
Wen, ZF ;
Saunders, R ;
Kee, ST ;
Sze, DY ;
Daniel, BL ;
Laerum, F ;
Pelc, NJ .
ACADEMIC RADIOLOGY, 2001, 8 (12) :1200-1207
[9]  
Fahrig R, 2001, J MAGN RESON IMAGING, V13, P294, DOI 10.1002/1522-2586(200102)13:2<294::AID-JMRI1042>3.3.CO
[10]  
2-O