Real-time three-dimensional motion analysis for patient positioning verification

被引:54
作者
Baroni, G
Ferrigno, G
Orecchia, R
Pedotti, A
机构
[1] Polytech Milan Univ, Dept Bioengn, I-20133 Milan, Italy
[2] Polytech Milan Univ, Ctr Bioengn, Pro Juventute Don Carlo Gnocchi Fdn, IRCCS, I-20148 Milan, Italy
[3] European Oncol Inst, Div Radiotherapy, IRCCS, I-20141 Milan, Italy
关键词
radiotherapy; quality control; patient positioning; real-time; three-dimensional motion analysis;
D O I
10.1016/S0167-8140(99)00166-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: This paper describes the technology and methods involved in a system for automatically checking the position of patients at radiotherapy units. This is proposed for improving the accuracy in the irradiation geometrical set-up, which is a crucial factor in radiotherapy quality control. Materials and methods: The system is based on real-time opto-electronics and close-range photogrammetry and detects multiple passive markers placed on selected patient skin landmarks. Patient alignment and position monitoring is carried our by comparing the current three-dimensional positions of the markers with those of an initial reference position acquired during the simulation procedure and/or the first irradiation session. The system was used to measure the accuracy of conventional laser centering techniques for patient repositioning. Inaccuracies due to breathing and random movements were also taken into account. Professional technicians were asked to reposition three volunteer subjects carefully using a traditional laser centering procedure. Results: The results revealed significant repositioning errors even in highly controlled conditions, affecting particularly body areas relatively far from the skin reference points used for laser alignment. Conclusions: The outcome of the experimental application of this technology confirms its potential as a tool for patient repositioning and automatic detection of any errors caused by breathing or other unpredictable movements. The real-time feedback on the patient's position given by the system provides operators with appropriate visual indices and allows them to take suitable countermeasures in case of significant failures, In addition, the use of the system output for automatic position control is envisaged. (C) 2000 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:21 / 27
页数:7
相关论文
共 24 条
[11]   PATTERN-RECOGNITION IN 3D AUTOMATIC HUMAN MOTION ANALYSIS [J].
FERRIGNO, G ;
BORGHESE, NA ;
PEDOTTI, A .
ISPRS JOURNAL OF PHOTOGRAMMETRY AND REMOTE SENSING, 1990, 45 (04) :227-246
[12]   3-DIMENSIONAL OPTICAL ANALYSIS OF CHEST-WALL MOTION [J].
FERRIGNO, G ;
CARNEVALI, P ;
ALIVERTI, A ;
MOLTENI, F ;
BEULCKE, G ;
PEDOTTI, A .
JOURNAL OF APPLIED PHYSIOLOGY, 1994, 77 (03) :1224-1231
[13]   ELITE - A DIGITAL DEDICATED HARDWARE SYSTEM FOR MOVEMENT ANALYSIS VIA REAL-TIME TV SIGNAL-PROCESSING [J].
FERRIGNO, G ;
PEDOTTI, A .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1985, 32 (11) :943-950
[14]  
Ferrigno G., 1995, 3 DIMENSIONAL ANAL H, P57
[15]  
Greco C, 1997, Radiol Med, V93, P254
[16]   Accuracy of patient positioning during radiotherapy for bladder and brain tumours [J].
Hanna, CL ;
Slade, S ;
Mason, MD ;
Burnet, NG .
CLINICAL ONCOLOGY, 1999, 11 (02) :93-98
[17]   CLINICAL USE OF ONLINE PORTAL IMAGING FOR DAILY PATIENT TREATMENT VERIFICATION [J].
HERMAN, MG ;
ABRAMS, RA ;
MAYER, RR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (04) :1017-1023
[18]   DETERMINING LOCATIONS OF INTRACEREBRAL LESIONS FOR PROTON RADIOTHERAPY [J].
LEVIN, CV ;
HOUGH, J ;
ADAMS, LP ;
BOONZAIER, D ;
RUTHER, H ;
WYNCHANK, S .
PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (10) :1393-1401
[19]  
Munro, 1995, Semin Radiat Oncol, V5, P115, DOI 10.1016/S1053-4296(95)80005-0
[20]  
Noel G, 1997, Cancer Radiother, V1, P249, DOI 10.1016/S1278-3218(97)89772-7