Digital radiography with a large-area, amorphous-silicon, flat-panel X-ray detector system

被引:59
作者
Spahn, M
Strotzer, M
Völk, M
Böhm, S
Geiger, B
Hahm, G
Feuerbach, S
机构
[1] Siemens AG, Med Engn Angio Fluoro & Rad Syst Dev, D-91301 Forchheim, Germany
[2] Univ Hosp Regensburg, Dept Diagnost Radiol, Regensburg, Germany
关键词
digital radiography; flat-panel detector; amorphous silicon; detective quantum efficiency;
D O I
10.1097/00004424-200004000-00007
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
RATIONALE AND OBJECTIVES. TO investigate the image quality of a digital radiography system with an amorphous-silicon, large-area, digital flat-panel detector. METHODS. A flat-panel detector based on a matrix of amorphous silicon was integrated into a projection radiography system, The scintillator consisted of a layer of structured cesium iodide, The active matrix size of 3000(2) pixels together with a pixel size of 143 mu m provided a large image area of 43 x 43 cm(2). Basic image quality parameters such as detective quantum efficiency (DQE) and modulation transfer function (MTF) were measured and compared with those obtained with conventional systems, RESULTS. The measurement of DQE yielded a high value of 70% at zero spatial frequency. At a system dose equivalent to 400 speed, the DQE of the digital system was a factor of two larger than the DQE of a storage phosphor or screen-him system within the entire spatial frequency range between zero and the Nyquist limit of 3.5 Line pairs per millimeter, The flat-panel detector furthermore has an MTF that is superior to that in regular screen-film systems and also provides a substantially larger dynamic range. CONCLUSIONS. This new technology demonstrates its potential to provide equal or superior image quality to conventional screen-him systems and to reduce patient exposure to radiation dose. The advantages of digital radiography systems, based on a hat-panel detector as an instant image display, facilitation of work flow in the radiology department, and digital networking and archiving, are well in sight.
引用
收藏
页码:260 / 266
页数:7
相关论文
共 13 条
[1]  
ANTONUK LE, 1993, PHYS MEDICA, V9, P63
[2]   Amorphous silicon x-ray image sensor [J].
Chabbal, J ;
Chaussat, C ;
Ducourant, T ;
Fritsch, L ;
Michailos, J ;
Spinnler, V ;
Vieux, G ;
Arques, M ;
Hahm, G ;
Hoheisel, M ;
Horbaschek, H ;
Schulz, R ;
Spahn, M .
PHYSICS OF MEDICAL IMAGING: MEDICAL IMAGING 1996, 1996, 2708 :499-510
[3]   New CsI/a-Si 17" x 17" X-ray flat panel detector provides superior detectivity and immediate direct digital output for General Radiography systems [J].
Chaussat, C ;
Chabbal, J ;
Ducourant, T ;
Spinnler, V ;
Vieux, G ;
Neyret, R .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :45-56
[4]   THE SOCIAL CONSTRUCTION OF THE GREAT-DEPRESSION - INDUSTRIAL-POLICY DURING THE 1930S IN THE UNITED-STATES, BRITAIN, AND FRANCE [J].
DOBBIN, FR .
THEORY AND SOCIETY, 1993, 22 (01) :1-56
[5]   Novel large area MIS-type X-ray image sensor for digital radiography [J].
Kameshima, T ;
Kaifu, N ;
Takami, E ;
Morishita, M ;
Yamazaki, T .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :453-462
[6]   Radiographic imaging characteristics of a direct conversion detector using selenium and thin film transistor array [J].
Lee, DL ;
Cheung, LK ;
Jeromin, LS ;
Palecki, EF ;
Rodricks, B .
PHYSICS OF MEDICAL IMAGING - MEDICAL IMAGING 1997, 1997, 3032 :88-96
[7]   Image quality of scintillator based X-ray electronic imagers [J].
Moy, JP .
PHYSICS OF MEDICAL IMAGING, 1998, 3336 :187-194
[8]   Self-normalizing method to measure the detective quantum efficiency of a wide range of x-ray detectors [J].
Stierstorfer, K ;
Spahn, M .
MEDICAL PHYSICS, 1999, 26 (07) :1312-1319
[9]   Clinical application of a flat-panel x-ray detector based on amorphous silicon technology:: Image quality and potential for radiation dose reduction in skeletal radiography [J].
Strotzer, M ;
Gmeinwieser, J ;
Völk, M ;
Fründ, R ;
Seitz, J ;
Manke, C ;
Albrich, H ;
Feuerbach, S .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1998, 171 (01) :23-27
[10]   Detection of simulated chest lesions with normal and reduced radiation dose -: Comparison of conventional screen-film radiography and a flat-panel x-ray detector based on amorphous silicon [J].
Strotzer, M ;
Gmeinwieser, JK ;
Völk, M ;
Fründ, R ;
Seitz, J ;
Feuerbach, S .
INVESTIGATIVE RADIOLOGY, 1998, 33 (02) :98-103