Automated sampling and imaging of analytes separated on thin-layer chromatography plates using desorption electrospray ionization mass spectrometry

被引:102
作者
Van Berkel, Gary J. [1 ]
Kertesz, Vilmos [1 ]
机构
[1] Oak Ridge Natl Lab, Div Chem Sci, Organ & Biol Mass Spectrometry Grp, Oak Ridge, TN 37831 USA
关键词
D O I
10.1021/ac060690a
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Modest modifications to the atmospheric sampling capillary of a commercial electrospray mass spectrometer and upgrades to an in-house-developed surface positioning control software package (HandsFree TLC/MS) were used to enable the automated sampling and imaging of analytes on and within large area surface substrates using desorption electrospray ionization mass spectrometry. Sampling and imaging of rhodamine dyes separated on TLC plates were used to illustrate some of the practical applications of this system. Examples are shown for user-defined spot sampling from separated bands on a TLC plate ( one or multiple spots), scanning of a complete development lane ( one or multiple lanes), or imaging of analyte bands in a development lane (i.e., multiple lane scans with close spacing). The post data acquisition processing and data display aspects of the software system are also discussed.
引用
收藏
页码:4938 / 4944
页数:7
相关论文
共 20 条
[1]   Self-aspirating atmospheric pressure chemical ionization source for direct sampling of analytes on surfaces and in liquid solutions [J].
Asano, KG ;
Ford, MJ ;
Tomkins, BA ;
Van Berkel, GJ .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2005, 19 (16) :2305-2312
[2]   Desorption electrospray ionization mass spectrometry for high-throughput analysis of pharmaceutical samples in the ambient environment [J].
Chen, HW ;
Talaty, NN ;
Takáts, Z ;
Cooks, RG .
ANALYTICAL CHEMISTRY, 2005, 77 (21) :6915-6927
[3]   Ambient mass spectrometry [J].
Cooks, RG ;
Ouyang, Z ;
Takats, Z ;
Wiseman, JM .
SCIENCE, 2006, 311 (5767) :1566-1570
[4]   Thin-layer chromatography/electrospray ionization triple-quadrupole linear ion trap mass spectrometry system:: analysis of rhodamine dyes separated on reversed-phase C8 plates [J].
Ford, MJ ;
Kertesz, V ;
Van Berkel, GJ .
JOURNAL OF MASS SPECTROMETRY, 2005, 40 (07) :866-875
[5]   Quantitative thin-layer chromatography/mass spectrometry analysis of caffeine using a surface sampling probe electrospray ionization tandem mass spectrometry system [J].
Ford, MJ ;
Deibel, MA ;
Tomkins, BA ;
Van Berkel, GJ .
ANALYTICAL CHEMISTRY, 2005, 77 (14) :4385-4389
[6]  
Ford MJ, 2004, RAPID COMMUN MASS SP, V18, P1303, DOI 10.1002/rcm.1486
[7]   SONIC SPRAY IONIZATION METHOD FOR ATMOSPHERIC-PRESSURE IONIZATION MASS-SPECTROMETRY [J].
HIRABAYASHI, A ;
SAKAIRI, M ;
KOIZUMI, H .
ANALYTICAL CHEMISTRY, 1994, 66 (24) :4557-4559
[8]   SONIC SPRAY MASS-SPECTROMETRY [J].
HIRABAYASHI, A ;
SAKAIRI, M ;
KOIZUMI, H .
ANALYTICAL CHEMISTRY, 1995, 67 (17) :2878-2882
[9]   Desorption electrospray ionization mass spectrometry for the analysis of pharmaceuticals and metabolites [J].
Kauppila, TJ ;
Wiseman, JM ;
Ketola, RA ;
Kotiaho, T ;
Cooks, RG ;
Kostiainen, R .
RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2006, 20 (03) :387-392
[10]   Automation of a surface sampling probe/electrospray mass spectrometry system [J].
Kertesz, V ;
Ford, MJ ;
Van Berkel, GJ .
ANALYTICAL CHEMISTRY, 2005, 77 (22) :7183-7189