Comparison of two versions of the Vigileo-FloTrac™ system (1.03 and 1.07) for stroke volume estimation: a multicentre, blinded comparison with oesophageal Doppler measurements

被引:31
作者
Chatti, R. [6 ]
de Rudniki, S. [2 ]
Marque, S. [4 ]
Dumenil, A. S. [3 ]
Descorps-Declere, A. [3 ]
Cariou, A. [4 ]
Duranteau, J. [2 ]
Aout, M. [5 ]
Vicaut, E. [5 ]
Cholley, B. P. [1 ]
机构
[1] Univ Paris 05, Hop Europeen Georges Pompidou, AP HP, Fac Med,Serv Anesthesie Reanimat, F-75908 Paris 15, France
[2] Univ Paris 11, Hop Bicetre, AP HP, Dept Anesthesie Reanimat, Paris, France
[3] Univ Paris 11, Hop Antoine Beclere, AP HP, Dept Anesthesie Reanimat, Paris, France
[4] Univ Paris 05, Hop Cochin, AP HP, Serv Reanimat Med, Paris, France
[5] Univ Paris 07, Hop Fernand Widal, AP HP, Unite Rech Clin, Paris, France
[6] Hop Lariboisiere, AP HP, Dept Anesthesie Reanimat, F-75475 Paris, France
关键词
equipment; flowmeters; measurement techniques; flow velocity waveform; monitoring; cardiopulmonary; WAVE-FORM ANALYSIS; OUTPUT MONITORING-SYSTEM; CARDIAC-OUTPUT; THERMODILUTION TECHNIQUE; SURGERY; AGREEMENT; VALIDITY;
D O I
10.1093/bja/aep020
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Our aim was to evaluate the validity of stroke volume measurements obtained using the Vigileo-FloTrac (TM) system in comparison with those obtained using oesophageal Doppler considered as a reference. Prospective, multicentre study (four university hospitals), in which investigators were blinded to stroke volume values acquired simultaneously with the other technique. Two different versions of the Vigileo software (1.03 and 1.07) were studied and compared over two consecutive periods of time. Forty critically ill patients (three ICUs) and 20 high-risk surgical patients (one operating theatre) were studied over a 6-month period. Two hundred and forty paired stroke volume values obtained using the second version of the Vigileo (1.07) yielded better correlation and agreement (R=0.48, P < 0.001; bias=4 ml, limits of agreement: +/- 41 ml) than the 207 paired values obtained using version 1.03 (R=0.12, P=0.1; bias=1 ml, limits of agreement: +/- 75 ml). However, even with the second version, the percentage error in stroke volume measurement was 58%, a value still above the range considered clinically acceptable (30%). The precision of stroke volume estimation using Vigileo-FloTrac (TM) has improved with the second version of the software (1.07), but remains insufficient to allow the replacement of the reference technique in the population studied.
引用
收藏
页码:463 / 469
页数:7
相关论文
共 28 条
  • [1] Agreement between methods of measurement with multiple observations per individual
    Bland, J. Martin
    Altman, Douglas G.
    [J]. JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) : 571 - 582
  • [2] Cardiac output measured by a new arterial pressure waveform analysis method without calibration compared with thermodilution after cardiac surgery
    Breukers, Rose-Marieke B. G. E.
    Sepehrkhouy, Shahrzad
    Spiegelenberg, Stefan R.
    Groeneveld, A. B. Johan
    [J]. JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2007, 21 (05) : 632 - 635
  • [3] Clinical evaluation of the FloTrac/Nigileo™ system and two established continuous cardiac output monitoring devices in patients undergoing cardiac surgery
    Button, D.
    Weibel, L.
    Reuthebuch, O.
    Genoni, M.
    Zollinger, A.
    Hofer, C. K.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2007, 99 (03) : 329 - 336
  • [4] Comparison of FloTrac™ cardiac output monitoring system in patients undergoing coronary artery bypass grafting with pulmonary artery cardiac output measurements
    Cannesson, M.
    Attof, Y.
    Rosamel, P.
    Joseph, P.
    Bastien, O.
    Lehot, J.-J.
    [J]. EUROPEAN JOURNAL OF ANAESTHESIOLOGY, 2007, 24 (10) : 832 - 839
  • [5] Noninvasive cardiac output monitoring by aortic blood flow determination:: Evaluation of the Sometec Dynemo-3000 system
    Cariou, A
    Monchi, M
    Joly, LM
    Bellenfant, F
    Claessens, YE
    Thébert, D
    Brunet, F
    Dhainaut, JF
    [J]. CRITICAL CARE MEDICINE, 1998, 26 (12) : 2066 - 2072
  • [6] Esophageal Doppler: Noninvasive cardiac output monitor
    Cholley, BP
    Singer, M
    [J]. ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2003, 20 (08): : 763 - 769
  • [7] Performance of a minimally invasive uncalibrated cardiac output monitoring system (Flotrac™/Vigileo™) in haemodynamically unstable patients
    Compton, F. D.
    Zukunft, B.
    Hoffmann, C.
    Zidek, W.
    Schaefer, J. -H.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (04) : 451 - 456
  • [8] A meta-analysis of studies using bias and precision statistics to compare cardiac output measurement techniques
    Critchley, LAH
    Critchley, JAJH
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 1999, 15 (02) : 85 - 91
  • [9] The validity of trans-esophageal Doppler ultrasonography as a measure of cardiac output in critically ill adults
    Dark, PM
    Singer, M
    [J]. INTENSIVE CARE MEDICINE, 2004, 30 (11) : 2060 - 2066
  • [10] Stroke volume variation obtained with FloTrac/Vigileo™ fails to predict fluid responsiveness in coronary artery bypass graft patients
    de Waal, E. E. C.
    Rex, S.
    Kruitwagen, C. L. J. J.
    Kalkman, C. J.
    Buhre, W. F.
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2008, 100 (05) : 725 - 726