Isotope-ratio detection for gas chromatography

被引:214
作者
Sessions, Alex L. [1 ]
机构
[1] CALTECH, Div Geol & Planetary Sci, Pasadena, CA 91125 USA
关键词
combustion interface; compound-specific isotope analysis; GC-isotope-ratio mass spectrometry; isotope-ratio-monitoring; pyrolysis interface;
D O I
10.1002/jssc.200600002
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Instrumentation and methods exist for highly precise analyses of the stable-isotopic composition of organic compounds separated by GC. The general approach combines a conventional GC, a chemical reaction interface, and a specialized isotoperatio mass spectrometer (IRMS). Most existing GC hardware and methods are amenable to isotope-ratio detection. The interface continuously and quantitatively converts all organic matter, including column bleed, to a common molecular form for isotopic measurement. C and N are analyzed as CO2 and N-2, respectively, derived from combustion of analytes. H and O are analyzed as H-2 and CO produced by pyrolysis/reduction. IRMS instruments are optimized to provide intense, highly stable ion beams, with extremely high precision realized via a system of differential measurements in which ion currents for all major isotopologs are simultaneously monitored. Calibration to an internationally recognized scale is achieved through comparison of closely spaced sample and standard peaks. Such systems are capable of measuring C-13/C-12 ratios with a precision approaching 0.1 parts per thousand (for values reported in the standard delta notation), four orders of magnitude better than that typically achieved by conventional "organic" mass spectrometers. Detection limits to achieve this level of precision are typically < 1 nmol C (roughly 10 ng of a typical hydrocarbon) injected on-column. Achievable precision and detection limits are correspondingly higher for N, O, and H, in that order.
引用
收藏
页码:1946 / 1961
页数:16
相关论文
共 98 条
[61]   IMPROVEMENTS IN MASS SPECTROMETERS FOR THE MEASUREMENT OF SMALL DIFFERENCES IN ISOTOPE ABUNDANCE RATIOS [J].
MCKINNEY, CR ;
MCCREA, JM ;
EPSTEIN, S ;
ALLEN, HA ;
UREY, HC .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1950, 21 (08) :724-730
[62]   Applied gas chromatography coupled to isotope ratio mass spectrometry [J].
Meier-Augenstein, W .
JOURNAL OF CHROMATOGRAPHY A, 1999, 842 (1-2) :351-371
[63]  
MeierAugenstein W, 1997, RAPID COMMUN MASS SP, V11, P1775, DOI 10.1002/(SICI)1097-0231(19971030)11:16<1775::AID-RCM89>3.0.CO
[64]  
2-P
[65]  
MEIERAUGENSTEIN W, 2004, HDB STABLE ISOTOPE A, P153
[66]   PERFORMANCE AND OPTIMIZATION OF A COMBUSTION INTERFACE FOR ISOTOPE RATIO MONITORING GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
MERRITT, DA ;
FREEMAN, KH ;
RICCI, MP ;
STUDLEY, SA ;
HAYES, JM .
ANALYTICAL CHEMISTRY, 1995, 67 (14) :2461-2473
[67]   NITROGEN ISOTOPIC ANALYSES BY ISOTOPE-RATIO-MONITORING GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
MERRITT, DA ;
HAYES, JM .
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1994, 5 (05) :387-397
[68]   CARBON ISOTOPIC ANALYSIS OF ATMOSPHERIC METHANE BY ISOTOPE-RATIO-MONITORING GAS-CHROMATOGRAPHY MASS-SPECTROMETRY [J].
MERRITT, DA ;
HAYES, JM ;
MARIAS, DJD .
JOURNAL OF GEOPHYSICAL RESEARCH-ATMOSPHERES, 1995, 100 (D1) :1317-1326
[69]   FACTORS CONTROLLING PRECISION AND ACCURACY IN ISOTOPE-RATIO-MONITORING MASS-SPECTROMETRY [J].
MERRITT, DA ;
HAYES, JM .
ANALYTICAL CHEMISTRY, 1994, 66 (14) :2336-2347
[70]   ISOTOPE-RATIO-MONITORING GAS-CHROMATOGRAPHY MASS-SPECTROMETRY - METHODS FOR ISOTOPIC CALIBRATION [J].
MERRITT, DA ;
BRAND, WA ;
HAYES, JM .
ORGANIC GEOCHEMISTRY, 1994, 21 (6-7) :573-583