共 24 条
A soft and flexible biosensor using a phospholipid polymer for continuous glucose monitoring
被引:27
作者:
Chu, MingXing
[2
]
Kudo, Hiroyuki
[1
]
Shirai, Takayuki
[3
]
Miyajima, Kumiko
[1
]
Saito, Hirokazu
[1
,4
]
Morimoto, Nobuyuki
[5
,6
]
Yano, Kazuyoshi
[7
]
Iwasaki, Yasuhiko
[8
]
Akiyoshi, Kazunari
[5
]
Mitsubayashi, Kohji
[1
]
机构:
[1] Tokyo Med & Dent Univ, Dept Biomed Devices & Instrumentat, Inst Biomat & Bioengn, Chiyoda Ku, Tokyo 1010062, Japan
[2] Tokyo Med & Dent Univ, Grad Sch Med & Dent Sci, Chiyoda Ku, Tokyo 1010062, Japan
[3] Tokyo Univ Technol, Grad Sch, Tokyo 1920982, Japan
[4] Tokyo Natl Coll Technol, Dept Mech Engn, Tokyo 1930997, Japan
[5] Tokyo Med & Dent Univ, Dept Organ Mat, Inst Biomat & Bioengn, Chiyoda Ku, Tokyo 1010062, Japan
[6] Tohoku Univ, Dept Mat Proc, Grad Sch Engn, Sendai, Miyagi 9808579, Japan
[7] Tokyo Univ Technol, Sch Bion, Tokyo 1920982, Japan
[8] Kansai Univ, Dept Chem & Mat Engn, Fac Chem Mat & Bioengn, Suitasi, Osakafu 5648680, Japan
基金:
日本学术振兴会;
关键词:
Biosensor;
Phospholipid polymer;
Flexible;
Glucose sensor;
TEAR GLUCOSE;
SENSOR;
TISSUE;
ELECTRODES;
D O I:
10.1007/s10544-009-9300-1
中图分类号:
R318 [生物医学工程];
学科分类号:
0831 ;
摘要:
A flexible biosensor using a phospholipid polymer to immobilization of glucose oxidase (GOD) was fabricated and tested. At first, an enzyme membrane formed by immobilizing GOD onto a porous polytetrafluoroethylene (PTFE) membrane using the phospholipid polymer (2-methacryloyloxyethyl phosphorylcholine (MPC) copolymerized with 2-ethylhexylmethacrylate (EHMA) : PMEH) was evaluated. According to the result of amperometric measurement, average density of GOD to be immobilized was optimized to 38.9 units cm(-2). Temperature and pH dependences were also investigated. Then, a flexible glucose sensor was fabricated by immobilizing GOD onto a flexible hydrogen peroxide electrode using PMEH. The flexible glucose sensor showed a linear relationship between output currents and glucose concentration in 0.05-1.00 mmol L-1, with a correlation coefficient of 0.999. The calibration range covered the normal tear glucose level of 0.14-0.23 mmol L-1. This indicates that the flexible biosensor is considered to be useful for monitoring of glucose in tear fluids.
引用
收藏
页码:837 / 842
页数:6