Iterative Reconstruction Technique for Reducing Body Radiation Dose at CT: Feasibility Study

被引:587
作者
Hara, Amy K. [1 ]
Paden, Robert G. [1 ]
Silva, Alvin C. [1 ]
Kujak, Jennifer L. [1 ]
Lawder, Holly J. [1 ]
Pavlicek, William [1 ]
机构
[1] Mayo Clin Arizona, Dept Diagnost Radiol, Scottsdale, AZ 85259 USA
关键词
body CT; low-dose CT; radiation dose; COMPUTED-TOMOGRAPHY; MDCT COLONOGRAPHY; IMAGE QUALITY; REDUCTION; EXPOSURE; LESIONS;
D O I
10.2214/AJR.09.2397
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
OBJECTIVE. The purpose of this study was to evaluate the image noise, low-contrast resolution, image quality, and spatial resolution of adaptive statistical iterative reconstruction in low-dose body CT. MATERIALS AND METHODS. Adaptive statistical iterative reconstruction was used to scan the American College of Radiology phantom at the American College of Radiology reference value and at one-half that value (12.5 mGy). Test objects in low- and high-contrast and uniformity modules were evaluated. Low-dose CT with adaptive statistical iterative reconstruction was then tested on 12 patients (seven men, five women average age, 67.5 years) who had previously undergone routine-dose CT. Two radiologists blinded to scanning technique evaluated images of the same patients obtained with routine-dose CT and low-dose CT with and without adaptive statistical iterative reconstruction. Image noise, low-contrast resolution, image quality, and spatial resolution were graded on a scale of 1 (best) to 4 (worst). Quantitative noise measurements were made on clinical images. RESULTS. In the phantom, low- and high-contrast and uniformity assessments showed no significant difference between routine-dose imaging and low-dose CT with adaptive statistical iterative reconstruction. In patients, low-dose CT with adaptive statistical iterative reconstruction was associated with CT dose index reductions of 32-65% compared with routine imaging and had the least noise both quantitatively and qualitatively (p < 0.05). Low-dose CT with adaptive statistical iterative reconstruction and routine-dose CT had identical results for low-contrast resolution and nearly identical results for overall image quality (grade 2.1-2.2). Spatial resolution was better with routine-dose CT (p = 0.004). CONCLUSION. These preliminary results support body CT dose index reductions of 32-65% when adaptive statistical iterative reconstruction is used. Studies with larger statistical samples are needed to confirm these findings.
引用
收藏
页码:764 / 771
页数:8
相关论文
共 22 条
[1]
[Anonymous], 2007, 2007 CT MARK SUMM RE
[2]
Current concepts - Computed tomography - An increasing source of radiation exposure [J].
Brenner, David J. ;
Hall, Eric J. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 357 (22) :2277-2284
[3]
Feasibility of MDCT colonography in ultra-low-dose technique in the detection of colorectal lesions: Comparison with high-resolution video colonoscopy [J].
Cohnen, M ;
Vogt, C ;
Beck, A ;
Andersen, K ;
Heinen, W ;
vom Dahl, S ;
Aurich, V ;
Haeussinger, D ;
Moedder, U .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2004, 183 (05) :1355-1359
[4]
Epidemiology of Obesity in the Western Hemisphere [J].
Ford, Earl S. ;
Mokdad, Ali H. .
JOURNAL OF CLINICAL ENDOCRINOLOGY & METABOLISM, 2008, 93 (11) :S1-S8
[5]
Dose reduction and image quality in MDCT colonography using tube current modulation [J].
Graser, A. ;
Wintersperger, B. J. ;
Suess, C. ;
Reiser, M. F. ;
Becker, C. R. .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2006, 187 (03) :695-701
[6]
COMPUTERIZED TRANSVERSE AXIAL SCANNING (TOMOGRAPHY) .1. DESCRIPTION OF SYSTEM [J].
HOUNDFIELD, GN .
BRITISH JOURNAL OF RADIOLOGY, 1973, 46 (552) :1016-1022
[7]
Detection and characterization of lesions on low-radiation-dose abdominal CT images postprocessed with noise reduction filters [J].
Kalra, MK ;
Maher, MM ;
Blake, MA ;
Lucey, BC ;
Karau, K ;
Toth, TL ;
Avinash, G ;
Halpern, EF ;
Saini, S .
RADIOLOGY, 2004, 232 (03) :791-797
[8]
Radiation from "extra" images acquired with abdominal and/or pelvic CT: Effect of automatic tube current modulation [J].
Kalra, MK ;
Maher, MM ;
Toth, TL ;
Kamath, RS ;
Halpern, EF ;
Saini, S .
RADIOLOGY, 2004, 232 (02) :409-414
[9]
Strategies for CT radiation dose optimization [J].
Kalra, MK ;
Maher, MM ;
Toth, TL ;
Hamberg, LM ;
Blake, MA ;
Shephard, JA ;
Saini, S .
RADIOLOGY, 2004, 230 (03) :619-628
[10]
Low-dose CT of the abdomen: Evaluation of image improvement with use of noise reduction filters - Pilot study [J].
Kalra, MK ;
Maher, MM ;
Sahani, DV ;
Blake, MA ;
Hahn, PF ;
Avinash, GB ;
Toth, TL ;
Halpern, E ;
Saini, S .
RADIOLOGY, 2003, 228 (01) :251-256