Selenium-based digital radiography in the detection of bone lesions: Preliminary experience with experimentally created defects

被引:10
作者
Ludwig, K [1 ]
Link, TM [1 ]
Fiebich, M [1 ]
Renger, B [1 ]
Diederich, S [1 ]
Oelerich, M [1 ]
Lenzen, H [1 ]
Heindel, W [1 ]
机构
[1] Univ Munster, Dept Clin Radiol, D-48129 Munster, Germany
关键词
animals; bones; radiography; femur; abnormalities; comparative studies; digital; selenium detector; storage phosphor; selenium;
D O I
10.1148/radiology.216.1.r00jl08220
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To compare the diagnostic performance of selenium-based digital radiography with that of conventional screen-film radiography and storage phosphor radiography for the detection of bone lesions simulating osteolyses. MATERIALS AND METHODS: Artificial osseous lesions 1.0-3.0 mm in diameter were created in 80 of 160 predefined regions in 16 porcine femoral specimens. Specimens were enclosed in containers filled with paraffin to ensure accurate repositioning and to obtain an absorption condition comparable to that of a human extremity imaging was performed with a selenium-based digital radiography system, a conventional screen-film system, and a storage phosphor radiography system with an exposure identical to that used during clinical imaging. The presence of a lesion was assessed with a five-point confidence scale. Receiver operating characteristic (ROC) analysis was performed for a total of 1,440 observations (480 per modality), and diagnostic performance was estimated with the area under the ROC curve (A,). Differences in diagnostic performance were assessed with the paired Student t test. RESULTS: ROC analysis results showed A, values of 0.656 for selenium-based digital radiography, 0.679 for storage phosphor radiography, and 0.680 for conventional screen-film radiography. Differences between the three modalities were not significant (P =.60-.93). CONCLUSION: Image quality with selenium-based digital radiography was comparable to that with conventional screen-film radiography and storage phosphor radiography.
引用
收藏
页码:220 / 224
页数:5
相关论文
共 22 条
[1]   Chest radiographic image quality: Comparison of asymmetric screen-film, digital storage phosphor, and digital selenium drum systems - Preliminary study [J].
Beute, GH ;
Flynn, MJ ;
Eyler, WR ;
Samei, E ;
Spizarny, DL ;
Zylak, CJ .
RADIOGRAPHICS, 1998, 18 (03) :745-754
[2]   TECHNICAL EVALUATION OF A DIGITAL CHEST RADIOGRAPHY SYSTEM THAT USES A SELENIUM DETECTOR [J].
CHOTAS, HG ;
FLOYD, CE ;
RAVIN, CE .
RADIOLOGY, 1995, 195 (01) :264-270
[3]   SELENIUM-BASED DIGITAL RADIOGRAPHY OF THE CHEST - RADIOLOGISTS PREFERENCE COMPARED WITH FILM-SCREEN RADIOGRAPHS [J].
FLOYD, CE ;
BAKER, JA ;
CHOTAS, HG ;
DELONG, DM ;
RAVIN, CE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (06) :1353-1358
[4]   DIGITAL CHEST IMAGING WITH A SELENIUM DETECTOR - COMPARISON WITH CONVENTIONAL RADIOGRAPHY FOR VISUALIZATION OF SPECIFIC ANATOMIC REGIONS OF THE CHEST [J].
HEESEWIJK, HPM ;
NEITZEL, U ;
VANDERGRAAF, Y ;
DEVALOIS, JC ;
FELDBERG, MAM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1995, 165 (03) :535-540
[5]  
HILL AB, 1971, PRINCIPLES MED STATI, P143
[6]  
Ishida M., 1983, RADIOGRAPHICS, V3, P325
[7]  
Kehler M, 1996, ACTA RADIOL, V37, P332
[8]   A comprehensive physical image quality evaluation of a selenium based digital x-ray imaging system for thorax radiography [J].
Launders, JH ;
Kengyelics, SM ;
Cowen, AR .
MEDICAL PHYSICS, 1998, 25 (06) :986-997
[9]   ARTIFICIAL BONE EROSIONS - DETECTION WITH MAGNIFICATION RADIOGRAPHY VERSUS CONVENTIONAL HIGH-RESOLUTION RADIOGRAPHY [J].
LINK, TM ;
RUMMENY, EJ ;
LENZEN, H ;
REUTER, I ;
ROOS, N ;
PETERS, PE .
RADIOLOGY, 1994, 192 (03) :861-864
[10]   INVESTIGATION OF BASIC IMAGING PROPERTIES IN DIGITAL RADIOGRAPHY .4. EFFECT OF UNSHARP MASKING ON THE DETECTABILITY OF SIMPLE PATTERNS [J].
LOO, LND ;
DOI, K ;
METZ, CE .
MEDICAL PHYSICS, 1985, 12 (02) :209-214