Serial immunoassays in parallel on a microfluidic chip for monitoring hormone secretion from living cells

被引:99
作者
Dishinger, John F.
Kennedy, Robert T. [1 ]
机构
[1] Univ Michigan, Dept Chem, Ann Arbor, MI 48109 USA
[2] Univ Michigan, Dept Pharmacol, Ann Arbor, MI 48109 USA
关键词
D O I
10.1021/ac061425s
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A microfluidic chip that allows for the continuous monitoring of cellular secretions from multiple independent living samples was developed. Performance of the device was characterized through the analysis of insulin secretion from islets of Langerhans. The chip contained four individual channel networks, each capable of performing electrophoresis-based immunoassays of the perfusate from islets. In the networks, islets were housed in a chamber that was continuously perfused with pressure-driven biological media at 0.6 mu L min(-1). Electroosmosis was used to pull perfusate containing secreted insulin into 4-cm-long reaction channels where it mixed with fluorescein isothiocyanate-labeled insulin and anti-insulin antibody for 60 s. The reaction streams were sampled at 6.25-s intervals and analyzed in parallel using an on-chip capillary electrophoresis separation with laser-induced fluorescence detection by a scanning confocal microscope. The limit of detection for insulin was 10 nM. The device was used to complete over 1450 immunoassays of biological samples in less than 40 min, allowing the parallel monitoring of insulin release from four islets every 6.25 s.
引用
收藏
页码:947 / 954
页数:8
相关论文
共 39 条
[21]  
PAWLEY JB, 1994, HDB BIOL CONFOCAL MI, P459
[22]   SINGLE ISLET BETA-CELL STIMULATION BY NUTRIENTS - RELATIONSHIP BETWEEN PYRIDINE-NUCLEOTIDES, CYTOSOLIC CA2+ AND SECRETION [J].
PRALONG, WF ;
BARTLEY, C ;
WOLLHEIM, CB .
EMBO JOURNAL, 1990, 9 (01) :53-60
[23]   Microfluidic chip for continuous monitoring of hormone secretion from live cells using an electrophoresis-based immunoassay [J].
Roper, MG ;
Shackman, JG ;
Dahlgren, GM ;
Kennedy, RT .
ANALYTICAL CHEMISTRY, 2003, 75 (18) :4711-4717
[24]   Acousto-optical deflection-based whole channel scanning for microchip isoelectric focusing with laser-induced fluorescence detection [J].
Sanders, JC ;
Huang, ZL ;
Landers, JP .
LAB ON A CHIP, 2001, 1 (02) :167-172
[25]   Microfluidic electrophoresis chip coupled to microdialysis for in vivo monitoring of amino acid neurotransmitters [J].
Sandlin, ZD ;
Shou, MS ;
Shackman, JG ;
Kennedy, RT .
ANALYTICAL CHEMISTRY, 2005, 77 (23) :7702-7708
[26]   Perfusion and chemical monitoring of living cells on a microfluidic chip [J].
Shackman, JG ;
Dahlgren, GM ;
Peters, JL ;
Kennedy, RT .
LAB ON A CHIP, 2005, 5 (01) :56-63
[27]   High-throughput automated post-processing of separation data [J].
Shackman, JG ;
Watson, CJ ;
Kennedy, RT .
JOURNAL OF CHROMATOGRAPHY A, 2004, 1040 (02) :273-282
[28]   Radial capillary array electrophoresis microplate and scanner for high-performance nucleic acid analysis [J].
Shi, YN ;
Simpson, PC ;
Scherer, JR ;
Wexler, D ;
Skibola, C ;
Smith, MT ;
Mathies, RA .
ANALYTICAL CHEMISTRY, 1999, 71 (23) :5354-5361
[29]   The design and fabrication of three-chamber microscale cell culture analog devices with integrated dissolved oxygen sensors [J].
Sin, A ;
Chin, KC ;
Jamil, MF ;
Kostov, Y ;
Rao, G ;
Shuler, ML .
BIOTECHNOLOGY PROGRESS, 2004, 20 (01) :338-345
[30]   On line competitive immunoassay for insulin based on capillary electrophoresis with laser induced fluorescence detection [J].
Tao, L ;
Kennedy, RT .
ANALYTICAL CHEMISTRY, 1996, 68 (22) :3899-3906