Validation of a precision radiochromic film dosimetry system for quantitative two-dimensional imaging of acute exposure dose distributions

被引:98
作者
Dempsey, JF [1 ]
Low, DA [1 ]
Mutic, S [1 ]
Markman, J [1 ]
Kirov, AS [1 ]
Nussbaum, GH [1 ]
Williamson, JF [1 ]
机构
[1] Washington Univ, Sch Med, Mallinckrodt Inst Radiol, Radiat Oncol Ctr, St Louis, MO 63110 USA
关键词
radiochromic film dosimetry; GafChromic film model MD-55; HDR remote afterloading system; brachytherapy dosimetry; conformal therapy dosimetry; IMRT; Monte Carlo photon transport simulation;
D O I
10.1118/1.1290488
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present an evaluation of the precision and accuracy of image-based radiochromic film (RCF) dosimetry performed using a commercial RCF product (Gafchromic MD-55-2, Nuclear Associates, Inc.) and a commercial high-spatial resolution (100 mum pixel size) He-Ne scanning-laser film-digitizer (Personal Densitometer, Molecular Dynamics, Inc.) as an optical density (OD) imaging system. The precision and accuracy of this dosimetry system are evaluated by performing RCF imaging dosimetry in well characterized conformal external beam and brachytherapy high dose-rate (HDR) radiation fields. Benchmarking of image-based RCF dosimetry is necessary due to many potential errors inherent to RCF dosimetry including: a temperature-dependent time evolution of RCF dose response; nonuniform response of RCF; and optical-polarization artifacts. In addition, laser-densitometer imaging artifacts can produce systematic OD measurement errors as large as 35% in the presence of high OD gradients. We present a RCF exposure and readout protocol that was developed for the accurate dosimetry of high dose rate (HDR) radiation sources. This protocol follows and expands upon the guidelines set forth by the American Association of Physicists in Medicine (AAPM) Task Group 55 report. Particular attention is focused on the OD imaging system, a scanning-laser film digitizer, modified to eliminate OD artifacts that were not addressed in the AAPM Task Group 55 report. RCF precision using this technique was evaluated with films given uniform 6 MV x-ray doses between 1 and 200 Gy. RCF absolute dose accuracy using this technique was evaluated by comparing RCF measurements to small volume ionization chamber measurements for conformal external-beam sources and an experimentally validated Monte Carlo photon-transport simulation code for a Ir-192 brachytherapy source. Pixel-to-pixel standard deviations of uniformly irradiated films were less than 1% for doses between 10 and 150 Gy; between 1% and 5% for lower doses down to 1 Gy and 1% and 1.5% for higher doses up to 200 Gy. Pixel averaging to form 200-800 mum pixels reduces these standard deviations by a factor of 2 to 5. Comparisons of absolute dose show agreement within 1.5%-4% of dose benchmarks, consistent with a highly accurate dosimeter limited by its observed precision and the precision of the dose standards to which it is compared. These results provide a comprehensive benchmarking of RCF, enabling its use in the commissioning of novel HDR therapy sources. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00510-1].
引用
收藏
页码:2462 / 2475
页数:14
相关论文
共 88 条
[1]  
[Anonymous], 1983, MED PHYS, V10, P741
[2]  
BJARNGARD BE, 1990, MED PHYS, V17, P794, DOI 10.1118/1.596475
[3]   Extrapolated surface dose measurements with radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
MEDICAL PHYSICS, 1999, 26 (03) :485-488
[4]   Effects of read-out light sources and ambient light on radiochromic film [J].
Butson, MJ ;
Yu, PKN ;
Metcalfe, PE .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (08) :2407-2412
[5]  
Carolan M. G., 1997, Australasian Physical and Engineering Sciences in Medicine, V20, P183
[6]   Monte Carlo calculations and GafChromic film measurements for plugged collimator helmets of Leksell Gamma Knife unit [J].
Cheung, JYC ;
Yu, KN ;
Ho, RTK ;
Yu, CP .
MEDICAL PHYSICS, 1999, 26 (07) :1252-1256
[7]   The electron-dose distribution surrounding an Ir-192 wire brachytherapy source investigated using EGS4 simulations and GafChromic(TM) film [J].
Cheung, YC ;
Yu, PKN ;
Young, ECM ;
Chan, CL ;
Ng, MF ;
Tang, FNF ;
Wong, TPY .
APPLIED RADIATION AND ISOTOPES, 1997, 48 (07) :985-990
[8]   HIGH-SENSITIVITY GAFCHROMIC FILM DOSIMETRY FOR I-125 SEED [J].
CHIUTSAO, ST ;
DELAZERDA, A ;
LIN, J ;
KIM, JH .
MEDICAL PHYSICS, 1994, 21 (05) :651-657
[9]   A stereotactic radiation therapy device for retinoblastoma using a noncircular collimator and intensity filter [J].
Cormack, RA ;
Kooy, HM ;
Bellerive, MR ;
Loeffler, JS ;
Petersen, RA ;
Tarbell, NJ .
MEDICAL PHYSICS, 1998, 25 (08) :1438-1442
[10]   THE NEEDS FOR BRACHYTHERAPY SOURCE CALIBRATIONS IN THE UNITED-STATES [J].
COURSEY, BM ;
GOODMAN, LJ ;
HOPPES, DD ;
LOEVINGER, R ;
MCLAUGHLIN, WL ;
SOARES, CG ;
WEAVER, JT .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 312 (1-2) :246-250