The effects of target size and tissue density on the minimum margin required for random errors

被引:48
作者
Witte, MG [1 ]
van der Geer, J [1 ]
Schneider, C [1 ]
Lebesque, JV [1 ]
van Herk, M [1 ]
机构
[1] Antoni Van Leeuwenhoek Hosp, Netherlands Canc Inst, Dept Radiotherapy, Amsterdam, Netherlands
关键词
radiation therapy; treatment planning; margins; setup errors; tissue density; target size;
D O I
10.1118/1.1809991
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The minimum margins required to compensate for random geometric uncertainties in the delivery of radiotherapy treatment were determined for a spherical Clinical Target Volume, using an analytic model for the cumulative dose. Margins were calculated such that the minimum dose in the target would be no less than 95% of the prescribed dose for 90% of the patients. The dose distribution model incorporated two Gaussians, and could accurately represent realistic dose profiles for various target sizes in lung and water. It was found that variations in target size and tissue density lead to significant changes in the minimum margin required for random errors. The random error margin increased with tissue density, and decreased with target size. The required margins were similar for dose distributions of spherical and cylindrical symmetry. Significant dose outside the spherical high dose region, as could result from multiple incident beams, lead to an increased margin for the larger targets. We could confirm that the previously proposed margin of 0.7 times the standard deviation of the random errors is safe for standard deviations up to 5 mm, except for very small targets in dense material. (C) 2004 American Association of Physicists in Medicine.
引用
收藏
页码:3068 / 3079
页数:12
相关论文
共 22 条
[1]   MEASUREMENT OF PROSTATE MOVEMENT OVER THE COURSE OF ROUTINE RADIOTHERAPY USING IMPLANTED MARKERS [J].
BALTER, JM ;
SANDLER, HM ;
LAM, K ;
BREE, RL ;
LICHTER, AS ;
TENHAKEN, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01) :113-118
[2]   Target margins for random geometrical treatment uncertainties in conformal radiotherapy [J].
Bel, A ;
vanHerk, M ;
Lebesque, JV .
MEDICAL PHYSICS, 1996, 23 (09) :1537-1545
[3]   High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imaging [J].
Bel, A ;
Vos, PH ;
Rodrigus, PTR ;
Creutzberg, CL ;
Visser, AG ;
Stroom, JC ;
Lebesque, JV .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (02) :321-332
[4]   The effect of breathing and set-up errors on the cumulative dose to a lung tumor [J].
Engelsmann, M ;
Damen, EMF ;
De Jaeger, K ;
van Ingen, KM ;
Mijnheer, BJ .
RADIOTHERAPY AND ONCOLOGY, 2001, 60 (01) :95-105
[5]   Intra- and inter-observer variability in contouring prostate and seminal vesicles: implications for conformal treatment planning [J].
Fiorino, C ;
Reni, M ;
Bolognesi, A ;
Cattaneo, GM ;
Calandrino, R .
RADIOTHERAPY AND ONCOLOGY, 1998, 47 (03) :285-292
[6]  
Greer P B, 1998, Australas Radiol, V42, P207, DOI 10.1111/j.1440-1673.1998.tb00494.x
[7]   Measurement of patient positioning errors in three-dimensional conformal radiotherapy of the prostate [J].
Hanley, J ;
Lumley, MA ;
Mageras, GS ;
Sun, J ;
Zelefsky, MJ ;
Leibel, SA ;
Fuks, Z ;
Kutcher, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 37 (02) :435-444
[8]   THE EFFECT OF SETUP UNCERTAINTIES ON THE TREATMENT OF NASOPHARYNX CANCER [J].
HUNT, MA ;
KUTCHER, GJ ;
BURMAN, C ;
FASS, D ;
HARRISON, L ;
LEIBEL, S ;
FUKS, Z .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (02) :437-447
[9]   IMPLEMENTATION OF RANDOM POSITIONING ERROR IN COMPUTERIZED RADIATION TREATMENT PLANNING SYSTEMS AS A RESULT OF FRACTIONATION [J].
LEONG, J .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (03) :327-334
[10]   ALPHA BETA-VALUE AND THE IMPORTANCE OF SIZE OF DOSE PER FRACTION FOR LATE COMPLICATIONS IN THE SUPRAGLOTTIC LARYNX [J].
MACIEJEWSKI, B ;
TAYLOR, JMG ;
WITHERS, HR .
RADIOTHERAPY AND ONCOLOGY, 1986, 7 (04) :323-326