Spectral irradiance measurements of tungsten lamps with filter radiometers in the spectral range 290 nm to 900 nm

被引:34
作者
Kübarsepp, T
Kärhä, P
Manoocheri, F
Nevas, S
Ylianttila, L
Ikonen, E
机构
[1] Aalto Univ, Inst Met Res, FIN-02015 Helsinki, Finland
[2] Radiat & Nucl Safety Author STUK, Helsinki, Finland
关键词
D O I
10.1088/0026-1394/37/4/7
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A method of measuring the absolute spectral irradiance of quartz-halogen-tungsten lamps is described, based on the known responsivity of a filter radiometer, the components of which are separately characterized. The characterization is described for the wide wavelength range essential for deriving the spectrum of a lamp, from 260 nm to 950 nm. Novel methods of interpolation and measurement are implemented for the spectral responsivity of the filter radiometer. The combined standard uncertainty of spectral irradiance measurements is less than 1.4 parts in 10(2) from 290 nm to 320 nm (ultraviolet B) and 4 parts in 10(3) from 440 nm to 900 nm (visible to near-infrared). As an example, the derived spectral irradiances of two lamps measured at the Helsinki University of Technology (HUT, Finland) are presented and compared with the measurement results of the National Institute of Standards and Technology (NIST, USA) and the Physikalisch-Technische Bundesanstalt (PTB, Germany). The comparisons indicate that the HUT spectral irradiance scale is between those of the NIST and the PTB in the wavelength range 290 nm to 900 nm. The long-term reproducibility of the spectral irradiance measurements is also presented. Over a period of two years, the reproducibility appears to be better than 1 part in 10(2).
引用
收藏
页码:305 / 312
页数:8
相关论文
共 31 条
[1]  
BOIVIN LP, 1989, ABSOLUTE RADIOMETRY, P170
[2]  
*BUR INT POIDS MEA, 1997, REP WORK GROUP AIR 2
[3]   Realization of a photometric scale based on cryogenic radiometry [J].
Clare, JF ;
Nield, KM .
METROLOGIA, 1998, 35 (04) :251-254
[4]   The NIST detector-based luminous intensity scale [J].
Cromer, CL ;
Eppeldauer, G ;
Hardis, JE ;
Larason, TC ;
Ohno, Y ;
Parr, AC .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1996, 101 (02) :109-132
[5]   A NEW DETERMINATION OF THE EMISSIVITY OF TUNGSTEN RIBBON [J].
DEVOS, JC .
PHYSICA, 1954, 20 (09) :690-&
[6]   TRAP DETECTORS AND THEIR PROPERTIES [J].
FOX, NP .
METROLOGIA, 1991, 28 (03) :197-202
[7]   New basis for the Australian realization of the candela [J].
Gardner, JL ;
Butler, DJ ;
Atkinson, EG ;
Wilkinson, FJ .
METROLOGIA, 1998, 35 (04) :235-239
[8]   Realization of a scale of absolute spectral response using the national institute of standards and technology high-accuracy cryogenic radiometer [J].
Gentile, TR ;
Houston, JM ;
Cromer, CL .
APPLIED OPTICS, 1996, 35 (22) :4392-4403
[9]   Spectral reflectance of silicon photodiodes [J].
Haapalinna, A ;
Karha, P ;
Ikonen, E .
APPLIED OPTICS, 1998, 37 (04) :729-732
[10]   Radiometric realization of the candela with a trap detector [J].
Ikonen, E ;
Karha, P ;
Lassila, A ;
Manoochehri, F ;
Fagerlund, H ;
Liedquist, L .
METROLOGIA, 1996, 32 (06) :689-692