Multienzyme electrochemical array sensor for determination of phenols and pesticides

被引:52
作者
Solná, R
Sapelnikova, S
Skládal, P
Winther-Nielsen, M
Carlsson, C
Emnéus, J
Ruzgas, T
机构
[1] Masaryk Univ, Dept Biochem, CS-61137 Brno, Czech Republic
[2] Lund Univ, Dept Analyt Chem, S-22100 Lund, Sweden
[3] DHI, Water & Environm, DK-2970 Horsholm, Denmark
[4] ANOX, S-22647 Lund, Sweden
关键词
amperometric screen-printed biosensor; flow-injection analysis; steady state system; tyrosinase; peroxidase; cholinesterase; phenols; pesticides; wastewater;
D O I
10.1016/j.talanta.2004.07.005
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The screen-printed four-electrode system was used as the amperometric transducer for determination of phenols and pesticides using immobilised tyrosinase. peroxidase. acetylcholinesterase and butyrylcholinesterase. Acetylthiocholine chloride was chosen as substrate for cholinesterases to measure inhibition by pesticides, hydrogen peroxide served as co-substrate for peroxidase to measure phenols. The compatibility of hydrolases and oxidoreductases working in the same array was studied. The detection of p-cresol, catechol and phenol as well as of pesticides including carbaryl, heptenophos and fenitrothion was carried out in flow-through and steady state arrangements. In addition. the effects of heavy metals (CL2+, Cd2+, Fe3+), fluoride (NaF), benzene and dimethylsulphoxide on cholinesterase activities ere evaluated. It was demonstrated that electrodes modified with hydrolases and oxidoreductases can function in the same array. The achieved R.S.D. values obtained for the flow system were below 4% for the same sensor and less than 10% within a group of five sensors. For the steady state system, R.S.D.s were approximately twice higher. One assay was completed in less than 6 min. The limit of detection for catechol Using tyrosinase was equal to 0.35 and 1.7 muM in the flow and steady state systems, respectively. On the contrary, lower limits of detection for pesticides were achieved in the steady state system-carbaryl 26 W, heptenophos 14 nM and fenitrothion 0.58 muM. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:349 / 357
页数:9
相关论文
共 52 条
[1]   Development of tyrosinase biosensor based on pH-sensitive field-effect transistors for phenols determination in water solutions [J].
Anh, TM ;
Dzyadevych, SV ;
Soldatkin, AP ;
Chien, ND ;
Jaffrezic-Renault, N ;
Chovelon, JM .
TALANTA, 2002, 56 (04) :627-634
[2]   DETERMINATION OF PHENOL AND CHLORINATED PHENOLIC-COMPOUNDS BASED ON A PPO-BIOELECTRODE AND ITS INHIBITION [J].
BESOMBES, JL ;
COSNIER, S ;
LABBE, P ;
REVERDY, G .
ANALYTICAL LETTERS, 1995, 28 (03) :405-424
[3]   DETERMINATION OF PHENOLIC-COMPOUNDS IN SURFACE-WATER USING ONLINE LIQUID-CHROMATOGRAPHIC PRECOLUMN-BASED COLUMN-SWITCHING TECHNIQUES [J].
BROUWER, ER ;
BRINKMAN, UAT .
JOURNAL OF CHROMATOGRAPHY A, 1994, 678 (02) :223-231
[4]   Amperometric flow-injection determination of phenolic compounds at self-assembled monolayer-based tyrosinase biosensors [J].
Campuzano, S ;
Serra, B ;
Pedrero, M ;
de Villena, FJM ;
Pingarrón, JM .
ANALYTICA CHIMICA ACTA, 2003, 494 (1-2) :187-197
[5]   Disposable tyrosinase-peroxidase bi-enzyme sensor for amperometric detection of phenols [J].
Chang, SC ;
Rawson, K ;
McNeil, CJ .
BIOSENSORS & BIOELECTRONICS, 2002, 17 (11-12) :1015-1023
[6]   Detection of pesticides using an amperometric biosensor based on ferophthalocyanine chemically modified carbon paste electrode and immobilized bienzymatic system [J].
Ciucu, AA ;
Negulescu, C ;
Baldwin, RP .
BIOSENSORS & BIOELECTRONICS, 2003, 18 (2-3) :303-310
[7]   A comparison of amperometric screen-printed, carbon electrodes and their application to the analysis of phenolic compounds present in beers [J].
Cummings, EA ;
Linquette-Mailley, S ;
Mailley, P ;
Cosnier, S ;
Eggins, BR ;
McAdams, ET .
TALANTA, 2001, 55 (05) :1015-1027
[8]   SELECTIVE DETERMINATION OF PHENOLS IN WATER BY A 2-TRAP TANDEM EXTRACTION SYSTEM FOLLOWED BY LIQUID-CHROMATOGRAPHY [J].
DICORCIA, A ;
MARCHESE, S ;
SAMPERI, R .
JOURNAL OF CHROMATOGRAPHY, 1993, 642 (1-2) :175-184
[9]   MULTIRESIDUE METHOD FOR PESTICIDES IN DRINKING-WATER USING A GRAPHITIZED CARBON-BLACK CARTRIDGE EXTRACTION AND LIQUID-CHROMATOGRAPHIC ANALYSIS [J].
DICORCIA, A ;
MARCHETTI, M .
ANALYTICAL CHEMISTRY, 1991, 63 (06) :580-585
[10]   Effect of pH on direct electron transfer between graphite and horseradish peroxidase [J].
Ferapontova, E ;
Puganova, E .
JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2002, 518 (01) :20-26