共 12 条
Fiberoptic measurement of tendon forces is influenced by skin movement artifact
被引:14
作者:
Erdemir, A
Hamel, AJ
Piazza, SJ
Sharkey, NA
机构:
[1] Penn State Univ, Ctr Locomot Studies, University Pk, PA 16802 USA
[2] Penn State Univ, Dept Kinesiol, University Pk, PA 16802 USA
[3] Penn State Univ, Dept Mech & Nucl Engn, University Pk, PA 16802 USA
[4] Penn State Univ, Dept Orthopaed & Rehabil, Hershey, PA 17033 USA
关键词:
skin effect;
fiberoptics;
force transducer;
tendon force;
implantable;
D O I:
10.1016/S0021-9290(02)00414-1
中图分类号:
Q6 [生物物理学];
学科分类号:
071011 ;
摘要:
Fiberoptic cables have previously been used for tendon force measurements in vivo. To measure forces in the Achilles tendon, a cable is passed mediolaterally through the skin and tendon, transverse to the loading axis. As the tendon is loaded, its fibers compress the cable and modulate the intensity of transmitted light, which can be related to tendon force by an in situ calibration. The relative movement between skin and tendon at the cable entry and exit sites may cause error by bending the cable and thus altering transducer output. Cadaver simulations of walking were conducted to compare fiberoptic measurements of Achilles tendon forces to known loads applied to the tendon by actuators attached in series. Force measurement errors, which were high when the skin was intact (RMS errors 24-81% peak forces), decreased considerably after skin removal (RMS errors 10-33% peak forces). The fiberoptic transducer is a useful tool for measurement of tendon forces in situ under natural loading conditions when skin can be removed, but caution should be exercised during in vivo use of this technique or under circumstances where skin is in contact with the fiberoptic cable at the insertion and exit sites. (C) 2003 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:449 / 455
页数:7
相关论文