Influence of loading rate and cable migration on fiberoptic measurement of tendon force

被引:22
作者
Erdemir, A
Piazza, SJ
Sharkey, NA
机构
[1] Penn State Univ, Ctr Locomot Studies, University Pk, PA 16802 USA
[2] Penn State Univ, Dept Kinesiol, University Pk, PA 16802 USA
[3] Penn State Univ, Dept Mech & Nucl Engn, University Pk, PA 16802 USA
[4] Penn State Univ, Dept Orthopaed & Rehabil, Hershey, PA 17033 USA
关键词
fiberoptics; optical fiber; force transducer; tendon force; implantable;
D O I
10.1016/S0021-9290(02)00010-6
中图分类号
Q6 [生物物理学];
学科分类号
071011 [生物物理学];
摘要
Several investigators have recently used fiberoptic cables to measure tendon forces in situ. The technique may be subject to significant error due to cable migration and differences in the loading rates used for calibration and those experienced during measurement. This in vitro study examined the impact of these potential sources of error on transducer accuracy. A fiberoptic cable was passed perpendicular to the fibers of four Achilles tendons in the mediolateral direction and each specimen was cyclically loaded to 1000 N. The influence of loading rate on transducer Output was investigated by comparing results from tests conducted at 20, 200 and 1000 N/s. The effect of cable migration was examined by comparing the outputs obtained after displacing the cable one tendon width medially and laterally along its path in the tendon and then repeating the 200 N/s testing protocol. It was possible to obtain nonlinear specimen-specific relationships between the fiberoptic output and tendon force. Differences in loading rate resulted in root-mean-square (RMS) errors not larger than 17% maximum load. Hysteresis effects caused RMS errors smaller than 5% maximum load. Cable migration errors were less than 27%. The total RMS error due to the combined effects of loading rate difference and cable movement was less than 32%. Fiberoptic measurement of tendon force is attractive due to its low cost, easy implementation and comparable accuracy relative to other implantable force transducers. Although additional factors such as cable placement, edge artifacts due where the transducer exits the skin and non-uniform loading may also influence fiberoptic output, careful control of loading rate and transducer movement during calibration is imperative if maximum accuracy is to be achieved. (C) 2002 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:857 / 862
页数:6
相关论文
共 16 条
[1]
DIRECT INVIVO TENDON FORCE MEASUREMENT SYSTEM [J].
AN, KN ;
BERGLUND, L ;
COONEY, WP ;
CHAO, EYS ;
KOVACEVIC, N .
JOURNAL OF BIOMECHANICS, 1990, 23 (12) :1269-1271
[2]
Individual muscle contributions to the in vivo Achilles tendon force [J].
Arndt, AN ;
Komi, PV ;
Brüggemann, GP ;
Lukkariniemi, J .
CLINICAL BIOMECHANICS, 1998, 13 (07) :532-541
[3]
A low profile human tendon force transducer: The influence of tendon thickness on calibration [J].
Dennerlein, JT ;
Miller, JM ;
Mote, CD ;
Rempel, DM .
JOURNAL OF BIOMECHANICS, 1997, 30 (04) :395-397
[4]
In vivo human triceps surae and quadriceps femoris muscle function in a squat jump and counter movement jump [J].
Finni, T ;
Komi, PV ;
Lepola, V .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2000, 83 (4-5) :416-426
[5]
Achilles tendon loading during walking: application of a novel optic fiber technique [J].
Finni, T ;
Komi, PV ;
Lukkariniemi, J .
EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY AND OCCUPATIONAL PHYSIOLOGY, 1998, 77 (03) :289-291
[6]
Calibration and application of an intra-articular force transducer for the measurement of patellar tendon graft forces:: an in situ evaluation [J].
Fleming, BC ;
Good, L ;
Peura, GD ;
Beynnon, BD .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1999, 121 (04) :393-398
[7]
Factors influencing the output of an implantable force transducer [J].
Fleming, BC ;
Peura, GD ;
Beynnon, BD .
JOURNAL OF BIOMECHANICS, 2000, 33 (07) :889-893
[8]
Rate-independent characteristics of an arthroscopically implantable force probe in the human Achilles tendon [J].
Hall, GW ;
Crandall, JR ;
Carmines, DV ;
Hale, JE .
JOURNAL OF BIOMECHANICS, 1999, 32 (02) :203-207
[9]
Evaluation of the implantable force transducer for chronic tendon-force recordings [J].
Herzog, W ;
Archambault, JM ;
Leonard, TR ;
Nguyen, HK .
JOURNAL OF BIOMECHANICS, 1996, 29 (01) :103-109
[10]
In-situ calibration of the implantable force transducer [J].
Herzog, W ;
Hasler, EM ;
Leonard, TR .
JOURNAL OF BIOMECHANICS, 1996, 29 (12) :1649-1652