Accuracy of the one-point in vivo calibration of "wired" glucose oxidase electrodes implanted in jugular veins of rats in periods of rapid rise and decline of the glucose concentration

被引:55
作者
Schmidtke, DW [1 ]
Heller, A [1 ]
机构
[1] Univ Texas, Dept Chem Engn, Austin, TX 78712 USA
关键词
D O I
10.1021/ac970932u
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The hypothesis of the feasibility of one-point in vivo calibration of intravenously implanted glucose sensors during periods of rapid rise and decline of venous blood glucose concentration was tested, Miniature (5 x 10(-4) cm(2) mass transporting area) glucose electrodes with 10-90% response times <2 min, that did not consume oxygen, were implanted in jugular veins of systemically heparinized rats and used in 4-h experiments, during which the blood glucose concentration was amperometrically monitored. The glucose electrodes were made by electrically connecting ("wiring") reaction centers of glucose oxidase through an electron-conducting redox hydrogel to gold electrode surfaces. The redox polymer and enzyme constituting the electrode sensing layer were immobilized by cross-linking, and thus the electrodes had no diffusional and readily leached redox mediator. One hour after their implantation, the electrodes accurately tracked the blood glucose concentration when calibrated in vivo by a one-point calibration, when the glucose concentration was steady, when rising rapidly, and when declining steeply. For an assumed 2-min lag time, the sensor readings were well correlated with the true blood glucose concentrations, with linear regression analysis yielding a slope of 0.97 +/- 0.07 and an intercept (bias) of 0.3 +/- 0.3 mM. The correlation coefficient, r(2), was 0.949 +/- 0.020, and the percent difference through the 2-22 mM range was 1.9 +/- 1.0%. The results suggest that, in combination with understanding and modeling of transient physiological differences between the subcutaneous and the blood glucose concentrations, it will be possible to calibrate by one-point in vivo calibration subcutaneously implanted sensors, even while the glucose concentration changes rapidly.
引用
收藏
页码:2149 / 2155
页数:7
相关论文
共 48 条
[31]   INVIVO MOLECULAR SENSING IN DIABETES-MELLITUS - AN IMPLANTABLE GLUCOSE SENSOR WITH DIRECT ELECTRON-TRANSFER [J].
PICKUP, JC ;
SHAW, GW ;
CLAREMONT, DJ .
DIABETOLOGIA, 1989, 32 (03) :213-217
[32]   A GLUCOSE MONITORING-SYSTEM FOR ON LINE ESTIMATION IN MAN OF BLOOD-GLUCOSE CONCENTRATION USING A MINIATURIZED GLUCOSE SENSOR IMPLANTED IN THE SUBCUTANEOUS TISSUE AND A WEARABLE CONTROL UNIT [J].
POITOUT, V ;
MOATTISIRAT, D ;
REACH, G ;
ZHANG, Y ;
WILSON, GS ;
LEMONNIER, F ;
KLEIN, JC .
DIABETOLOGIA, 1993, 36 (07) :658-663
[33]   PHOTO-CROSS-LINKED COPOLYMERS OF 2-HYDROXYETHYL METHACRYLATE, POLY(ETHYLENE GLYCOL) TETRA-ACRYLATE AND ETHYLENE DIMETHACRYLATE FOR IMPROVING BIOCOMPATIBILITY OF BIOSENSORS [J].
QUINN, CP ;
PATHAK, CP ;
HELLER, A ;
HUBBELL, JA .
BIOMATERIALS, 1995, 16 (05) :389-396
[34]   Reverse iontophoresis: Noninvasive glucose monitoring in vivo in humans [J].
Rao, G ;
Guy, RH ;
Glikfeld, P ;
LaCourse, WR ;
Leung, L ;
Tamada, J ;
Potts, RO ;
Azimi, N .
PHARMACEUTICAL RESEARCH, 1995, 12 (12) :1869-1873
[35]  
REBRIN K, 1989, DIABETOLOGIA, V32, P573
[36]   GLUCOSE-CONCENTRATION IN SUBCUTANEOUS EXTRACELLULAR-SPACE [J].
SCHMIDT, FJ ;
SLUITER, WJ ;
SCHOONEN, AJM .
DIABETES CARE, 1993, 16 (05) :695-700
[37]   CALIBRATION OF A WEARABLE GLUCOSE SENSOR [J].
SCHMIDT, FJ ;
AALDERS, AL ;
SCHOONEN, AJM ;
DOORENBOS, H .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1992, 15 (01) :55-61
[38]   Statistics for critical clinical decision making based on readings of pairs of implanted sensors [J].
Schmidtke, DW ;
Pishko, MV ;
Quinn, CP ;
Heller, A .
ANALYTICAL CHEMISTRY, 1996, 68 (17) :2845-2849
[39]   Measurement and modeling of the transient difference between blood and subcutaneous glucose concentrations in the rat after injection of insulin [J].
Schmidtke, DW ;
Freeland, AC ;
Heller, A ;
Bonnecaze, RT .
PROCEEDINGS OF THE NATIONAL ACADEMY OF SCIENCES OF THE UNITED STATES OF AMERICA, 1998, 95 (01) :294-299
[40]  
SHICHIRI M, 1983, DIABETOLOGIA, V24, P179