Parallel acquisition techniques in cardiac cine magnetic resonance imaging using TrueFISP sequences: Comparison of image quality and artifacts

被引:44
作者
Hunold, P
Maderwald, S
Ladd, ME
Jellus, V
Barkhausen, J
机构
[1] Univ Hosp, Dept Diagnost & Intervent Radiol, D-45122 Essen, Germany
[2] Ruhr Univ Bochum, RF & Microwave Engn Inst, D-4630 Bochum, Germany
[3] Siemens Med Solut, Erlangen, Germany
关键词
cardiovascular MRI; steady-state free precession; clue MRI; reconstruction algorithms; parallel acquisition techniques; image quality;
D O I
10.1002/jmri.20125
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare image quality, artifacts, and signal-to-noise ratio (SNR) in cardiac cine TrueFISP magnetic resonance imaging (MRI) with and without parallel acquisition techniques (PAT). Materials and Methods: MRI was performed in 16 subjects with a TrueFISP sequence (1.5 T; Magnetom Sonata, Siemens): TR, 3.0 msec; TE, 1.5 msec; flip angle (FA), 60degrees. Three axes were scanned without PAT (no PAT) and using the generalized autocalibrating partially parallel acquisition (GRAPPA) and modified sensitivity encoding (mSENSE) reconstruction algorithms with an autocalibration mode to reduce scan time. A conventional spine array and a body flex array were used. Artifacts, image noise, and overall image quality were classified on a 4-point scale by an observer blinded to the implemented technique; for quantitative comparison, SNR was measured. Results: With a PAT factor of two, acquisition time could be reduced by 39%. No PAT did not show artifacts, and GRAPPA revealed fewer artifacts than mSENSE. PAT provided inferior-quality scores concerning image noise and overall image quality. In quantitative measurements, GRAPPA and mSENSE (20.1 +/- 6.2 and 15.6 +/- 6.2, respectively) yielded lower SNR than no PAT (30.6 +/- 20.1; P < 0.05) and P < 0.001). Conclusion: Time savings in PAT are accompanied by artifacts and an increase in image noise. The GRAPPA algorithm was superior to mSENSE concerning image quality, noise, and SNR.
引用
收藏
页码:506 / 511
页数:6
相关论文
共 20 条
[1]   MR evaluation of ventricular function: True fast imaging with steady-state precession versus fast low-angle shot cine MR imaging: Feasibility study [J].
Barkhausen, J ;
Ruehm, SG ;
Goyen, M ;
Buck, T ;
Laub, G ;
Debatin, JF .
RADIOLOGY, 2001, 219 (01) :264-269
[2]   Quantification of right and left ventricular function by cardiovascular magnetic resonance [J].
Bellenger, NG ;
Grothues, F ;
Smith, GC ;
Pennell, DJ .
HERZ, 2000, 25 (04) :392-399
[3]   Magnetic resonance of coronary arteries [J].
Bunce, NH ;
Pennell, DJ .
EUROPEAN RADIOLOGY, 2001, 11 (05) :721-731
[4]   AN ALGORITHM FOR NMR IMAGING RECONSTRUCTION BASED ON MULTIPLE RF RECEIVER COILS [J].
CARLSON, JW .
JOURNAL OF MAGNETIC RESONANCE, 1987, 74 (02) :376-380
[5]   Cine MR angiography of the heart with segmented true fast imaging with steady-state precession [J].
Carr, JC ;
Simonetti, O ;
Bundy, J ;
Li, DB ;
Pereles, S ;
Finn, JP .
RADIOLOGY, 2001, 219 (03) :828-834
[6]   A comparison of two methods for measuring the signal to noise ratio on MR images [J].
Firbank, MJ ;
Coulthard, A ;
Harrison, RM ;
Williams, ED .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (12) :N261-N264
[7]  
Griswold MA, 2000, MAGN RESON MATER PHY, V10, P105
[8]   VD-AUTO-SMASH imaging [J].
Heidemann, RM ;
Griswold, MA ;
Haase, A ;
Jakob, PM .
MAGNETIC RESONANCE IN MEDICINE, 2001, 45 (06) :1066-1074
[9]  
Jakob PM, 2000, MAGN RESON MATER PHY, V11, P52, DOI 10.1016/S1352-8661(00)00112-5
[10]   Accelerated cardiac imaging using the SMASH technique [J].
Jakob, PM ;
Griswold, MA ;
Edelman, RR ;
Manning, WJ ;
Sodickson, DK .
JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 1999, 1 (02) :153-157