A fully scalable matrix-based magnetoresistive biochip has been recently introduced for the detection of biomolecular recognition events. This work reports results for a 3 x 3 cells matrix comprising hydrogenated amorphous silicon diodes (PIN and Schottky) and aluminum oxide magnetic tunnel junctions (MTJ). The noise characteristics of the individual thin film diodes (TFD) and the MTJ were analyzed for measuring frequencies from DC to 1 kHz, allowing the detection of single magnetic nanoparticles with 250 mn and 130 nm diameters. For 50 nm particles, and at 1 kHz, minimum number of detectable labels ranges from 20 (Schottky diodes) to 50 (PIN diodes).
引用
收藏
页码:2403 / 2405
页数:3
相关论文
共 12 条
[1]   Diode/magnetic tunnel junction cell for fully scalable matrix-based biochip [J].
Cardoso, F. A. ;
Ferreira, H. A. ;
Conde, J. P. ;
Chu, V. ;
Freitas, P. P. ;
Vidal, D. ;
Germano, J. ;
Sousa, L. ;
Piedade, M. S. ;
Costa, B. A. ;
Lemos, J. M. .
JOURNAL OF APPLIED PHYSICS, 2006, 99 (08)
[2]   Ion beam deposition and oxidation of spin-dependent tunnel junctions [J].
Cardoso, S ;
Gehanno, V ;
Ferreira, R ;
Freitas, PP .
IEEE TRANSACTIONS ON MAGNETICS, 1999, 35 (05) :2952-2954
[3]   Planar Hall effect sensor for magnetic micro- and nanobead detection [J].
Ejsing, L ;
Hansen, MF ;
Menon, AK ;
Ferreira, HA ;
Graham, DL ;
Freitas, PP .
APPLIED PHYSICS LETTERS, 2004, 84 (23) :4729-4731
[4]   Rapid DNA hybridization based on ac field focusing of magnetically labeled target DNA [J].
Ferreira, HA ;
Feliciano, N ;
Graham, DL ;
Clarke, LA ;
Amaral, MD ;
Freitas, PP .
APPLIED PHYSICS LETTERS, 2005, 87 (01)
[5]   Detection of cystic fibrosis related DNA targets using AC field focusing of magnetic labels and spin-valve sensors [J].
Ferreira, HA ;
Graham, DL ;
Feliciano, N ;
Clarke, LA ;
Amaral, MD ;
Freitas, PP .
IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (10) :4140-4142
[6]   Tuning of MgO barrier magnetic tunnel junction bias current for picotesla magnetic field detection [J].
Ferreira, R ;
Wisniowski, P ;
Freitas, PP ;
Langer, J ;
Ocker, B ;
Maass, W .
JOURNAL OF APPLIED PHYSICS, 2006, 99 (08)
[7]  
Freitas P.P., 2004, Magneto Electronics
[8]   Magnetoresistive-based biosensors and biochips [J].
Graham, DL ;
Ferreira, HA ;
Freitas, PP .
TRENDS IN BIOTECHNOLOGY, 2004, 22 (09) :455-462
[9]   Design and performance of GMR sensors for the detection of magnetic microbeads in biosensors [J].
Rife, JC ;
Miller, MM ;
Sheehan, PE ;
Tamanaha, CR ;
Tondra, M ;
Whitman, LJ .
SENSORS AND ACTUATORS A-PHYSICAL, 2003, 107 (03) :209-218
[10]   Comparison of a prototype magnetoresistive biosensor to standard fluorescent DNA detection [J].
Schotter, J ;
Kamp, PB ;
Becker, A ;
Pühler, A ;
Reiss, G ;
Brückl, H .
BIOSENSORS & BIOELECTRONICS, 2004, 19 (10) :1149-1156