AN AXIAL VELOCITY ESTIMATOR FOR ULTRASOUND BLOOD-FLOW IMAGING, BASED ON A FULL EVALUATION OF THE DOPPLER EQUATION BY MEANS OF A 2-DIMENSIONAL AUTOCORRELATION APPROACH

被引:403
作者
LOUPAS, T [1 ]
POWERS, JT [1 ]
GILL, RW [1 ]
机构
[1] ADV TECHNOL LABS, BOTHELL, WA 98041 USA
关键词
D O I
10.1109/58.393110
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
This paper introduces a new velocity estimator, referred to as the 2D autocorrelator, which differs from conventional Doppler techniques in two respects: the derivation of axial velocity values by evaluating the Doppler equation using explicit estimates of both the mean Doppler and the mean RF frequency at each range gate location; and, the 2D nature (depth samples versus pulse transmissions) of processing within the range gate, The estimator's output can be calculated by evaluating the 2D autocorrelation function of the demodulated (baseband) backscattered echoes at two lags, A full derivation and mathematical description of the estimator is presented, based on the framework of the 2D Fourier transform, The same framework is adopted to analyze two other established velocity estimators (the conventional 1D autoeorrelator and the crosscorrelator) in a unifying manner, and theoretical arguments as well as experimental results are used to highlight the common aspects of all three estimators, In addition, a thorough performance evaluation is carried out by means of extensive simulations, which document the effect of a number of factors (velocity spread, range gate length, ensemble length, noise level, transmitted bandwidth) and provide an insight into the optimum parameters and trade-offs associated with individual algorithms. Overall, the 2D autocorrelator is shown to offer the best performance in the context of the specific simulation conditions considered here, Its superiority over the crosscorrelator is restricted to cases of low signal-to-noise ratios, However, the 2D autocorrelator always outperforms the conventional 1D autocorrelator by a significant margin, These comparisons, when linked to the computational requirements of the proposed estimator, suggest that it combines the generally higher performance of 2D broadband time-domain techniques with the relatively modest complexity of 1D narrowband phase-domain velocity estimators.
引用
收藏
页码:672 / 688
页数:17
相关论文
共 49 条
[21]   A REAL-TIME ULTRASOUND TIME-DOMAIN CORRELATION BLOOD FLOWMETER .2. PERFORMANCE AND EXPERIMENTAL-VERIFICATION [J].
HEIN, IA ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1993, 40 (06) :776-785
[22]   A REAL-TIME ULTRASOUND TIME-DOMAIN CORRELATION BLOOD FLOWMETER .1. THEORY AND DESIGN [J].
HEIN, IA ;
CHEN, JT ;
JENKINS, WK ;
OBRIEN, WD .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1993, 40 (06) :768-775
[23]  
Herman G.T., 1980, IMAGE RECONSTRUCTION
[24]   ADAPTIVE ESTIMATION OF THE MEAN FREQUENCY OF A DOPPLER SIGNAL FROM SHORT DATA WINDOWS [J].
HERMENT, A ;
DEMOMENT, G ;
GUGLIELMI, JP ;
DUMEE, P ;
PELLOT, C .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (09) :901-919
[25]   COMPARISON OF THE PERFORMANCE OF THE RF CROSS-CORRELATION AND DOPPLER AUTOCORRELATION TECHNIQUE TO ESTIMATE THE MEAN VELOCITY OF SIMULATED ULTRASOUND SIGNALS [J].
HOEKS, APG ;
ARTS, TGJ ;
BRANDS, PJ ;
RENEMAN, RS .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (09) :727-740
[26]   TIME-DELAY ESTIMATION VIA CROSS-CORRELATION IN THE PRESENCE OF LARGE ESTIMATION ERRORS [J].
IANNIELLO, JP .
IEEE TRANSACTIONS ON ACOUSTICS SPEECH AND SIGNAL PROCESSING, 1982, 30 (06) :998-1003
[27]   DISCRETE-TIME TECHNIQUES FOR TIME-DELAY ESTIMATION [J].
JACOVITTI, G ;
SCARANO, G .
IEEE TRANSACTIONS ON SIGNAL PROCESSING, 1993, 41 (02) :525-533
[28]   RANGE VELOCITY LIMITATIONS FOR TIME-DOMAIN BLOOD VELOCITY ESTIMATION [J].
JENSEN, JA .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1993, 19 (09) :741-749
[29]   REAL-TIME TWO-DIMENSIONAL BLOOD-FLOW IMAGING USING AN AUTO-CORRELATION TECHNIQUE [J].
KASAI, C ;
NAMEKAWA, K ;
KOYANO, A ;
OMOTO, R .
IEEE TRANSACTIONS ON SONICS AND ULTRASONICS, 1985, 32 (03) :458-464