Accuracy and precision of time-averaged flow as measured by nontriggered 2D phase-contrast MR angiography, a phantom evaluation

被引:83
作者
Bakker, CJG
Kouwenhoven, M
Hartkamp, MJ
Hoogeveen, RM
Mali, WPTM
机构
[1] Department of Radiology, University Hospital Utrecht, 3584CX Utrecht
[2] Philips Medical Systems, Best
[3] Computer Vision Research Group, University of Utrecht
关键词
magnetic resonance; magnetic resonance angiography; flow studies; flow dynamics; pulsatile flow; cardiac triggering; phantom studies;
D O I
10.1016/0730-725X(95)02005-E
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this study was to assess the accuracy and precision of time-averaged flow as measured by nontriggered 2D PC. Mono-, bi-, and triphasic flow patterns, modelling waveforms encountered in the human vascular system, were generated by a computer-controlled flow system. Time-averaged flow velocity was measured by conventional 2D cardiac-triggered cine PC and by nontriggered 2D PC for different settings of the excitation flip angle and the velocity sensitivity. Accuracy and precision were determined by repeating the measurements (N = 6) and comparing the results against precisely known calibration values. Measurements revealed waveform-specific deviations between triggered and nontriggered acquisitions that depended on the velocity sensitivity and, more strongly, on the flip angle of the nontriggered experiment. This confirmed the theoretically predicted predominance of amplitude over phase effects. Systematic errors could be reduced by decreasing the flip angle and the velocity sensitivity, although at the expense of signal-to-noise, so that additional signal averaging was required to maintain a specified precision. The attainable accuracy appeared to be acceptable only for waveforms with a relatively low pulsatility index. The study demonstrates the feasibility of accurate and precise nontriggered velocity measurements for weakly pulsatile flow and indicates a route towards improving the reliability for highly pulsatile flow.
引用
收藏
页码:959 / 965
页数:7
相关论文
共 13 条
  • [1] [Anonymous], 1969, DATA REDUCTION ERROR
  • [2] FACTORS INFLUENCING THE ACCURACY AND PRECISION OF VELOCITY-ENCODED PHASE IMAGING
    BUONOCORE, MH
    BOGREN, H
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 26 (01) : 141 - 154
  • [3] CORRELATION OF CINE MR VELOCITY-MEASUREMENTS IN THE INTERNAL CAROTID-ARTERY WITH COLLATERAL FLOW IN THE CIRCLE OF WILLIS - PRELIMINARY-STUDY
    DAVIS, WL
    TURSKI, PA
    GORBATENKO, KG
    WEBER, D
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (04): : 603 - 609
  • [4] ENZMANN DR, 1994, AM J NEURORADIOL, V15, P123
  • [5] COMPARISON OF CEREBRAL-ARTERY BLOOD-FLOW MEASUREMENTS WITH GATED CINE AND UNGATED PHASE-CONTRAST TECHNIQUES
    ENZMANN, DR
    MARKS, MP
    PELC, NJ
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (05): : 705 - 712
  • [6] A GEOMETRICALLY ACCURATE VASCULAR PHANTOM FOR COMPARATIVE-STUDIES OF X-RAY, ULTRASOUND, AND MAGNETIC-RESONANCE VASCULAR IMAGING - CONSTRUCTION AND GEOMETRICAL VERIFICATION
    FRAYNE, R
    GOWMAN, LM
    RICKEY, DW
    HOLDSWORTH, DW
    PICOT, PA
    DRANGOVA, M
    CHU, KC
    CALDWELL, CB
    FENSTER, A
    RUTT, BK
    [J]. MEDICAL PHYSICS, 1993, 20 (02) : 415 - 425
  • [7] COMPUTER-CONTROLLED FLOW SIMULATOR FOR MR FLOW STUDIES
    FRAYNE, R
    HOLDSWORTH, DW
    GOWMAN, LM
    RICKEY, DW
    DRANGOVA, M
    FENSTER, A
    RUTT, BK
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1992, 2 (05): : 605 - 612
  • [8] ANALYSIS OF MR PHASE-CONTRAST MEASUREMENTS OF PULSATILE VELOCITY WAVE-FORMS
    HANGIANDREOU, NJ
    ROSSMAN, PJ
    RIEDERER, SJ
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1993, 3 (02): : 387 - 394
  • [9] NONTRIGGERED MAGNETIC-RESONANCE VELOCITY-MEASUREMENT OF THE TIME-AVERAGE OF PULSATILE VELOCITY
    HOFMAN, MBM
    KOUWENHOVEN, M
    SPRENGER, M
    VANROSSUM, AC
    VALK, J
    WESTERHOF, N
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1993, 29 (05) : 648 - 655
  • [10] DETERMINATION OF CEREBRAL BLOOD-FLOW WITH A PHASE-CONTRAST CINE MR IMAGING TECHNIQUE - EVALUATION OF NORMAL SUBJECTS AND PATIENTS WITH ARTERIOVENOUS-MALFORMATIONS
    MARKS, MP
    PELC, NJ
    ROSS, MR
    ENZMANN, DR
    [J]. RADIOLOGY, 1992, 182 (02) : 467 - 476