Fat attenuation using a dual steady-state balanced-SSFP sequence with periodically variable flip angles

被引:18
作者
Absil, J
Denolin, V
Metens, T
机构
[1] Free Univ Brussels, Hop Erasme, Unite IRM Radiol, B-1070 Brussels, Belgium
[2] Univ Libre Bruxelles, Fac Sci, CENOLI, Brussels, Belgium
关键词
steady-state free precession; fast imaging; variable flip angles; fat suppression; off-resonance profile;
D O I
10.1002/mrm.20766
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A refocused-SSFP sequence based on balanced-FFE (TrueFisp, Fiesta) that attenuates fat signal is presented. The sequence uses periodically variable flip angles and produces a dual steady state of the signal, which is obtained after a dual transient phase if an appropriate preparation is used. The off-resonance profile of the steady-state signal exhibits large stopbands that can be employed for fat suppression. Numerical simulations were performed to investigate the signal behavior and the off-resonance properties of the sequence. Experimental results obtained with a Philips Gyroscan Intera 1.5T MR scanner demonstrated fat attenuation in phantoms and abdominal images in volunteers.
引用
收藏
页码:343 / 351
页数:9
相关论文
共 19 条
[1]   STEADY-STATE FREE PRECESSION IN NUCLEAR MAGNETIC RESONANCE [J].
CARR, HY .
PHYSICAL REVIEW, 1958, 112 (05) :1693-1701
[2]  
Deimling M., 1994, PROC 2 ANN M SMR SAN, P495
[3]   Reduction of transient signal oscillations in True-FISP using a linear flip angle series magnetization preparation [J].
Deshpande, VS ;
Chung, YC ;
Zhang, Q ;
Shea, SM ;
Li, DB .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) :151-157
[4]   SIMPLE PROTON SPECTROSCOPIC IMAGING [J].
DIXON, WT .
RADIOLOGY, 1984, 153 (01) :189-194
[5]   Fat-suppressed steady-state free precession imaging using phase detection [J].
Hargreaves, BA ;
Vasanawala, SS ;
Nayak, KS ;
Hu, BS ;
Nishimura, DG .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (01) :210-213
[6]   Characterization and reduction of the transient response in steady-state MR imaging [J].
Hargreaves, BA ;
Vasanawala, SS ;
Pauly, JM ;
Nishimura, DG .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (01) :149-158
[7]   Fat and water separation in balanced steady-state free precession using the Dixon method [J].
Huang, TY ;
Chung, HW ;
Wang, FN ;
Ko, CW ;
Chen, CY .
MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (02) :243-247
[8]   Steady-state sequence synthesis and its application to efficient fat-suppressed imaging [J].
Overall, WR ;
Nishimura, DG ;
Hu, BS .
MAGNETIC RESONANCE IN MEDICINE, 2003, 50 (03) :550-559
[10]   Rapid imaging cartilage with precession and of articular steady-state free multipoint fat water separation [J].
Reeder, SB ;
Pelc, NJ ;
Alley, MT ;
Gold, GE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 180 (02) :357-362