Evaluation of respiratory movement during gated radiotherapy using film and electronic portal imaging

被引:206
作者
Ford, EC
Mageras, GS
Yorke, E
Rosenzweig, KE
Wagman, R
Ling, CC
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Phys Med, New York, NY 10021 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10021 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2002年 / 52卷 / 02期
关键词
gating; respiratory motion; electronic portal imaging device; fluoroscopic images;
D O I
10.1016/S0360-3016(01)02681-5
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To evaluate the effectiveness of a commercial system in reducing respiration-induced treatment uncertainty by gating the radiation delivery. Methods and Materials: The gating system considered here measures respiration from the position of a reflective marker on the patient's chest. Respiration-triggered planning CT scans were obtained for 8 patients (4 lung, 4 liver) at the intended phase of respiration (6 at end expiration and 2 at end inspiration). In addition, fluoroscopic movies were recorded simultaneously with the respiratory waveform. During the treatment sessions, gated localization films were used to measure the position of the diaphragm relative to the vertebral bodies, which was compared to the reference digitally reconstructed radiograph derived from the respiration-triggered planning CT. Variability was quantified by the standard deviation about the mean position. We also assessed the interfraction variability of soft tissue structures during gated treatment in 2 patients using an amorphous silicon electronic portal imaging device. Results: The gated localization films revealed an interfraction patient-averaged diaphragm variability of 2.8 +/- 1.0 mm (error bars indicate standard deviation in the patient population). The fluoroscopic data yielded a patient-averaged intrafraction diaphragm variability of 2.6 +/- 1.7 mm. With no gating, this intrafraction excursion became 6.9 +/- 2.1 mm. In gated localization films, the patient-averaged mean displacement of the diaphragm from the planning position was 0.0 +/- 3.9 mm. However, in 4 of the 8 patients, the mean (over localization films) displacement was >4 mm, indicating a systematic displacement in treatment position from the planned one. The position of soft tissue features observed in portal images during gated treatments over several fractions showed a mean variability between 2.6 and 5.7 mm. The intrafraction variability, however, was between 0.6 and 1.4 mm, indicating that most of the variability was due to patient setup errors rather than to respiratory motion. Conclusions: The gating system evaluated here reduces the intra- and interfraction variability of anatomy due to respiratory motion. However, systematic displacements were observed in some cases between the location of an anatomic feature at simulation and its location during treatment. Frequent monitoring is advisable with film or portal imaging. (C) 2002 Elsevier Science Inc.
引用
收藏
页码:522 / 531
页数:10
相关论文
共 30 条
[1]   Determination of ventilatory liver movement via radiographic evaluation of diaphragm position [J].
Balter, JM ;
Dawson, LA ;
Kazanjian, S ;
McGinn, C ;
Brock, KK ;
Lawrence, T ;
Ten Haken, R .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (01) :267-270
[2]   High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imaging [J].
Bel, A ;
Vos, PH ;
Rodrigus, PTR ;
Creutzberg, CL ;
Visser, AG ;
Stroom, JC ;
Lebesque, JV .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (02) :321-332
[3]   Considerations for the implementation of target volume protocols in radiation therapy [J].
Craig, T ;
Battista, J ;
Moiseenko, V ;
Van Dyk, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (01) :241-250
[4]   ULTRASOUND QUANTITATION OF RESPIRATORY ORGAN MOTION IN THE UPPER ABDOMEN [J].
DAVIES, SC ;
HILL, AL ;
HOLMES, RB ;
HALLIWELL, M ;
JACKSON, PC .
BRITISH JOURNAL OF RADIOLOGY, 1994, 67 (803) :1096-1102
[5]   The reproducibility of organ position using active breathing control (ABC) during liver radiotherapy [J].
Dawson, LA ;
Brock, KK ;
Kazanjian, S ;
Fitch, D ;
McGinn, CJ ;
Lawrence, TS ;
Ten Haken, RK ;
Balter, J .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 51 (05) :1410-1421
[6]   What margins should be added to the clinical target volume in radiotherapy treatment planning for lung cancer? [J].
Ekberg, L ;
Holmberg, O ;
Wittgren, L ;
Bjelkengren, G ;
Landberg, T .
RADIOTHERAPY AND ONCOLOGY, 1998, 48 (01) :71-77
[7]   TIME TREND OF PATIENT SETUP DEVIATIONS DURING PELVIC IRRADIATION USING ELECTRONIC PORTAL IMAGING [J].
ELGAYED, AAH ;
BEL, A ;
VIJLBRIEF, R ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) :162-171
[8]   Deep inspiration breath-hold technique for lung tumors: The potential value of target immobilization and reduced lung density in dose escalation [J].
Hanley, J ;
Debois, MM ;
Mah, D ;
Mageras, GS ;
Raben, A ;
Rosenzweig, K ;
Mychalczak, B ;
Schwartz, LH ;
Gloeggler, PJ ;
Lutz, W ;
Ling, CC ;
Leibel, SA ;
Fuks, Z ;
Kutcher, GJ .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 45 (03) :603-611
[9]   Extracranial stereotactic radiation therapy:: Set-up accuracy of patients treated for liver metastases [J].
Herfarth, KK ;
Debus, J ;
Lohr, F ;
Bahner, ML ;
Fritz, P ;
Höss, A ;
Schlegel, W ;
Wannenmacher, MF .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 46 (02) :329-335
[10]  
Keatley EL, 2000, USE OF COMPUTERS IN RADIATION THERAPY, P132