High-resolution dose profiles produced by the Leksell Gamma Knife were obtained in BANG(TM) polymer gel, using a 3 T whole-body scanner upgraded by a magnetic resonance microscopy unit. The gel was contained in 22.3 mm diameter flasks that were inserted into a solid, tissue-equivalent head phantom irradiated by fields of by 8 and 14 mm collimators. Dose profiles were obtained from a linear dose-response curve (R-2 vs. Dose). Excellent agreement was obtained when the gel data were compared to film dosimetry and calculated data. (C) 2000 Elsevier Science Inc. All rights reserved.
引用
收藏
页码:343 / 349
页数:7
相关论文
共 31 条
[1]   ABSORBED DOSE PERTURBATION CAUSED BY DIODES FOR SMALL-FIELD PHOTON DOSIMETRY [J].
BEDDAR, AS ;
MASON, DJ ;
OBRIEN, PF .
MEDICAL PHYSICS, 1994, 21 (07) :1075-1079
[2]   ERRORS IN T2 ESTIMATION USING MULTISLICE MULTIPLE-ECHO IMAGING [J].
CRAWLEY, AP ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1987, 4 (01) :34-47
[3]   A novel 675.2 nm diode laser densitometer for use with GafChromic films [J].
Ertl, A ;
Kellermann, PO ;
Zehetmayer, M ;
Schöggl, A ;
Kindl, P ;
Maitz, AH .
MEDICAL PHYSICS, 1999, 26 (05) :834-838
[4]   Shuttle dose at the Vienna Leksell Gamma Knife [J].
Ertl, A ;
Zehetmayer, M ;
Schoggl, A ;
Kitz, K ;
Koschuch, R ;
Stadtmann, H ;
Kindl, P .
PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (06) :1567-1578
[5]   PROPERTIES OF THE PHASE-ALTERNATING PHASE-SHIFT (PHAPS) MULTIPLE SPIN-ECHO PROTOCOL IN MRI - A STUDY OF THE EFFECTS OF IMPERFECT RF PULSES [J].
FRANSSON, A ;
ERICSSON, A ;
JUNG, B ;
SPERBER, GO .
MAGNETIC RESONANCE IMAGING, 1993, 11 (06) :771-784
[6]   Test objects for MRI quality assurance based on polymer gels [J].
Gore, JC ;
Maryanski, MJ ;
Schulz, RJ .
MEDICAL PHYSICS, 1997, 24 (09) :1405-1408
[7]   RADIATION-DAMAGE INDUCED DOSE-RATE NON-LINEARITY IN AN N-TYPE SILICON DETECTOR [J].
GRUSELL, E ;
RIKNER, G .
ACTA RADIOLOGICA ONCOLOGY, 1984, 23 (06) :465-469
[8]   A comparison of dosimetry techniques in stereotactic radiosurgery [J].
Heydarian, M ;
Hoban, PW ;
Beddoe, AH .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (01) :93-110
[9]   Three-dimensional visualization and measurement of conformal dose distributions using magnetic resonance imaging of BANG polymer gel dosimeters [J].
Ibbott, GS ;
Maryanski, MJ ;
Eastman, P ;
Holcomb, SD ;
Zhang, YS ;
Avison, RG ;
Sanders, M ;
Gore, JC .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (05) :1097-1103
[10]   Dosimeter gel and MR imaging for verification of calculated dose distributions in clinical radiation therapy [J].
Johansson, SA ;
Magnusson, P ;
Fransson, A ;
Olsson, LE ;
Christoffersson, JO ;
Montelius, A ;
Mattsson, S .
ACTA ONCOLOGICA, 1997, 36 (03) :283-290