Continuous ASL (CASL) perfusion MRI with an array coil and parallel imaging at 3T

被引:67
作者
Wang, Z
Wang, JJ
Connick, TJ
Wetmore, GS
Detre, JA
机构
[1] Univ Penn, Sch Med, Ctr Funct Neuroimaging, Dept Neurol, Philadelphia, PA 19104 USA
[2] Univ Penn, Sch Med, Dept Radiol, Philadelphia, PA 19104 USA
关键词
ASL perfusion imaging; parallel imaging; GRAPPA; CBF; SNR;
D O I
10.1002/mrm.20574
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The purpose of this work was to assess the feasibility and efficacy of using an array coil and parallel imaging in continuous arterial spin labeling (CASL) perfusion MRI. An 8-channel receive-only array head coil was used in conjunction with a surrounding detunable volume transmit coil. The signal to noise ratio (SNR), temporal stability, cerebral blood flow (CBF), and perfusion image coverage were measured from steady state CASL scans using: a standard volume coil, array coil, and array coil with 2- and 3-fold accelerated parallel imaging. Compared to the standard volume coil, the array coil provided 3 times the average SNR increase and higher temporal stability for the perfusion weighted images, even with threefold acceleration. Although perfusion images of the array coil were affected by the inhomogeneous coil sensitivities, this effect was invisible in the quantitative CBF images, which showed highly reproducible perfusion values compared to the standard volume coil. The unfolding distortions of parallel imaging were suppressed in the perfusion images by pairwise subtraction, though they sharply degraded the raw EPI images. Moreover, parallel imaging provided the potential of acquiring more slices due to the shortened acquisition time and improved coverage in brain regions with high static field inhomogeneity. Such results highlight the potential utility of array coils and parallel imaging in ASL perfusion MRI.
引用
收藏
页码:732 / 737
页数:6
相关论文
共 20 条
  • [1] Multisection cerebral blood flow MR imaging with continuous arterial spin labeling
    Alsop, DC
    Detre, JA
    [J]. RADIOLOGY, 1998, 208 (02) : 410 - 416
  • [2] Reduced transit-time sensitivity in noninvasive magnetic resonance imaging of human cerebral blood flow
    Alsop, DC
    Detre, JA
    [J]. JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1996, 16 (06) : 1236 - 1249
  • [3] Application of sensitivity-encoded echo-planar imaging for blood oxygen level-dependent functional brain imaging
    de Zwart, JA
    van Gelderen, P
    Kellman, P
    Duyn, JH
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2002, 48 (06) : 1011 - 1020
  • [4] Design of a SENSE-optimized high-sensitivity MRI receive coil for brain imaging
    de Zwart, JA
    Ledden, PJ
    Kellman, P
    van Gelderen, P
    Duyn, JH
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) : 1218 - 1227
  • [5] PERFUSION IMAGING
    DETRE, JA
    LEIGH, JS
    WILLIAMS, DS
    KORETSKY, AP
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1992, 23 (01) : 37 - 45
  • [6] A phased array echoplanar imaging system for fMRI
    Frederick, BD
    Wald, LL
    Maas, LC
    Renshaw, PF
    [J]. MAGNETIC RESONANCE IMAGING, 1999, 17 (01) : 121 - 129
  • [7] Generalized Autocalibrating Partially Parallel Acquisitions (GRAPPA)
    Griswold, MA
    Jakob, PM
    Heidemann, RM
    Nittka, M
    Jellus, V
    Wang, JM
    Kiefer, B
    Haase, A
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2002, 47 (06) : 1202 - 1210
  • [8] Haacke EM, 1999, MAGNETIC RESONANCE I
  • [9] Physiological noise in oxygenation-sensitive magnetic resonance imaging
    Krüger, G
    Glover, GH
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (04) : 631 - 637
  • [10] High resolution T2-weighted imaging of the human brain using surface coils and an analytical reception profile correction
    Moyher, SE
    Wald, LL
    Nelson, SJ
    Hallam, D
    Dillon, WP
    Norman, D
    Vigneron, DB
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (03) : 512 - 517