Computer-assisted thoracic pedicle screw placement -: An in vitro feasibility study

被引:75
作者
Kim, KD
Johnson, JP
Bloch, O
Masciopinto, JE
机构
[1] Univ Calif Davis, Dept Neurol Surg, Sacramento, CA 95817 USA
[2] Univ Calif Los Angeles, Div Neurosurg, Los Angeles, CA USA
[3] Rocky Mt Neurosurg Alliance PC, Englewood, CO USA
关键词
computer-assistance; image guidance; thoracic pedicle fixation;
D O I
10.1097/00007632-200102150-00011
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design: In this cadaveric study, a computer-assisted image guidance system was tested for accuracy of thoracic pedicle screw placement. Objectives: Evaluate the system's accuracy for thoracic pedicle screw placement in vitro. Summary of Background Data: The effective use and reliability of pedicle screw instrumentation in providing short-segment stabilization and correction of deformity is well known in the lumbar spine. Pedicle screw placement in the thoracic spine is difficult because of the small dimensions of the thoracic pedicles and risk to the adjacent spinal cord and neurovascular structures. Investigators have shown the improved accuracy of computer-assisted lumbar pedicle screw placement; but the accuracy of computer-assisted thoracic pedicle screw placement, which is becoming more widely used, has not been shown, Methods: In five human cadavers, 120 thoracic pedicle Screws were placed with computer-assisted image guidance. The largest clinically feasible screw was used based on the cross-sectional dimensions of each pedicle. The accuracy was assessed by postoperative computed tomography and visual inspection. Results: The overall pedicle cortex violation was 23 of 120 pedicles (19.2%). Nine violations (7.5%) were graded as major and 14 (11.7%) as minor. A marked and progressive learning curve was evident with the perforation rates that decreased from 37.5% in the first cadaver to 4.2% in the last two cadavers. Conclusions: Accurate thoracic pedicle screw placement is feasible with computer-assisted surgery. However, as with any other new surgical technology, the learning curve must be recognized and incorporated into the necessary fundamental knowledge and experience for these procedures.
引用
收藏
页码:360 / 364
页数:5
相关论文
共 24 条
[1]   COMPUTER-ASSISTED PEDICLE SCREW FIXATION - A FEASIBILITY STUDY [J].
AMIOT, LP ;
LABELLE, H ;
DEGUISE, JA ;
SATI, M ;
BRODEUR, P ;
RIVARD, CH .
SPINE, 1995, 20 (10) :1208-1212
[2]  
AN HS, 1991, SPINE S, V16, P548
[3]   In vitro simulation - Early results of stereotaxy for pedicle screw placement [J].
Carl, AL ;
Khanuja, HS ;
Sachs, BL ;
Gatto, CA ;
vomLehn, J ;
Vosburgh, K ;
Schenck, J ;
Lorensen, W ;
Rohling, K ;
Disler, D .
SPINE, 1997, 22 (10) :1160-1164
[4]   Projection of the thoracic pedicle and its morphometric analysis [J].
Ebraheim, NA ;
Xu, RM ;
Ahmad, M ;
Yeasting, RA .
SPINE, 1997, 22 (03) :233-238
[5]   Anatomic relations of the thoracic pedicle to the adjacent neural structures [J].
Ebraheim, NA ;
Jabaly, G ;
Xu, RM ;
Yeasting, RA .
SPINE, 1997, 22 (14) :1553-1556
[6]   Computer-aided pedicle screw placement using frameless stereotaxis [J].
Glossop, ND ;
Hu, RW ;
Randle, JA .
SPINE, 1996, 21 (17) :2026-2034
[7]   APPLICATION OF FRAMELESS STEREOTAXY TO PEDICLE SCREW FIXATION OF THE SPINE [J].
KALFAS, IH ;
KORMOS, DW ;
MURPHY, MA ;
MCKENZIE, RL ;
BARNETT, GH ;
BELL, GR ;
STEINER, CP ;
TRIMBLE, MB ;
WEISENBERGER, JP .
JOURNAL OF NEUROSURGERY, 1995, 83 (04) :641-647
[8]   Technical accuracy of a neuronavigation system measured with a high-precision mechanical micromanipulator [J].
Kaus, M ;
Steinmeier, R ;
Sporer, T ;
Ganslandt, O ;
Fahlbusch, R .
NEUROSURGERY, 1997, 41 (06) :1431-1436
[9]   Universal calibration of surgical instruments for spinal stereotaxy [J].
Kim, KD ;
Johnson, JP ;
Masciopinto, JE ;
Bloch, O ;
Saracen, MJ ;
Villablanca, JP .
NEUROSURGERY, 1999, 44 (01) :173-177
[10]   Improved accuracy of pedicle screw insertion with computer-assisted surgery - A prospective clinical trial of 30 patients [J].
Laine, T ;
Schlenzka, D ;
Makitalo, K ;
Tallroth, K ;
Nolte, LP ;
Visarius, H .
SPINE, 1997, 22 (11) :1254-1258