Position-probability-sampled Monte Carlo calculation of VMAT, 3DCRT, step-shoot IMRT, and helical tomotherapy dose distributions using BEAMnrc/DOSXYZnrc

被引:15
作者
Belec, Jason [1 ,2 ]
Ploquin, Nicolas [1 ]
La Russa, Daniel J. [1 ]
Clark, Brenda G. [1 ,2 ]
机构
[1] Ottawa Hosp Canc Ctr, Dept Med Phys, Ottawa, ON K1H 8L6, Canada
[2] Carleton Univ, Ottawa, ON K1S 5B6, Canada
关键词
Monte Carlo; intensity modulated arc therapy; helical tomotherapy; MULTILEAF COLLIMATOR; COMPONENT MODULE; VERIFICATION; SIMULATION; MODEL; BEAM; OPTIMIZATION; ACCELERATOR; VALIDATION; DELIVERY;
D O I
10.1118/1.3538922
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
100231 [临床病理学]; 100902 [航空航天医学];
摘要
Purpose: The commercial release of volumetric modulated arc therapy techniques using a conventional linear accelerator and the growing number of helical tomotherapy users have triggered renewed interest in dose verification methods, and also in tools for exploring the impact of machine tolerance and patient motion on dose distributions without the need to approximate time-varying parameters such as gantry position, MLC leaf motion, or patient motion. To this end we have developed a Monte Carlo-based calculation method capable of simulating a wide variety of treatment techniques without the need to resort to discretization approximations. Methods: The ability to perform complete position-probability-sampled Monte Carlo dose calculations was implemented in the BEAMnrc/DOSXZYnrc user codes of EGSnrc. The method includes full accelerator head simulations of our tomotherapy and Elekta linacs, and a realistic representation of continous motion via the sampling of a time variable. The functionality of this algorithm was tested via comparisons with both measurements and treatment planning dose distributions for four types of treatment techniques: 3D conformal, step-shoot intensity modulated radiation therapy, helical tomotherapy, and volumetric modulated arc therapy. Results: For static fields, the absolute dose agreement between the EGSnrc Monte Carlo calculations and measurements is within 2%/1 mm. Absolute dose agreement between Monte Carlo calculations and treatment planning system for the four different treatment techniques is within 3%/3 mm. Discrepancies with the tomotherapy TPS on the order of 10%/5 mm were observed for the extreme example of a small target located 15 cm off-axis and planned with a low modulation factor. The increase in simulation time associated with using position-probability sampling, as opposed to the discretization approach, was less than 2% in most cases. Conclusions: A single Monte Carlo simulation method can be used to calculate patient dose distribution for various types of treatment techniques delivered with either tomotherapy or a conventional linac. The method simplifies the simulation process, improves dose calculation accuracy, and involves an acceptably small change in computation time. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3538922]
引用
收藏
页码:948 / 960
页数:13
相关论文
共 30 条
[1]
GEANT4-a simulation toolkit [J].
Agostinelli, S ;
Allison, J ;
Amako, K ;
Apostolakis, J ;
Araujo, H ;
Arce, P ;
Asai, M ;
Axen, D ;
Banerjee, S ;
Barrand, G ;
Behner, F ;
Bellagamba, L ;
Boudreau, J ;
Broglia, L ;
Brunengo, A ;
Burkhardt, H ;
Chauvie, S ;
Chuma, J ;
Chytracek, R ;
Cooperman, G ;
Cosmo, G ;
Degtyarenko, P ;
Dell'Acqua, A ;
Depaola, G ;
Dietrich, D ;
Enami, R ;
Feliciello, A ;
Ferguson, C ;
Fesefeldt, H ;
Folger, G ;
Foppiano, F ;
Forti, A ;
Garelli, S ;
Giani, S ;
Giannitrapani, R ;
Gibin, D ;
Cadenas, JJG ;
González, I ;
Abril, GG ;
Greeniaus, G ;
Greiner, W ;
Grichine, V ;
Grossheim, A ;
Guatelli, S ;
Gumplinger, P ;
Hamatsu, R ;
Hashimoto, K ;
Hasui, H ;
Heikkinen, A ;
Howard, A .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 506 (03) :250-303
[2]
Multileaf collimator interleaf transmission [J].
Balog, JP ;
Mackie, TR ;
Wenman, DL ;
Glass, M ;
Fang, G ;
Pearson, D .
MEDICAL PHYSICS, 1999, 26 (02) :176-186
[3]
Briesmeister J.F., 2000, MCNP-A general Monte Carlo N -particle transport code, version 4A, V2
[4]
A technique for generating phase-space-based Monte Carlo beamlets in radiotherapy applications [J].
Bush, K. ;
Popescu, I. A. ;
Zavgorodni, S. .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (18) :N337-N347
[5]
Comparison of dose calculation algorithms with Monte Carlo methods for photon arcs [J].
Chow, JCL ;
Wong, E ;
Chen, JZ ;
Van Dyk, J .
MEDICAL PHYSICS, 2003, 30 (10) :2686-2694
[6]
Dose measurements compared with Monte Carlo simulations of narrow 6 MV multileaf collimator shaped photon beams [J].
De Vlamynck, K ;
Palmans, H ;
Verhaegen, F ;
De Wagter, C ;
De Neve, W ;
Thierens, H .
MEDICAL PHYSICS, 1999, 26 (09) :1874-1882
[7]
CERR: A computational environment for radiotherapy research [J].
Deasy, JO ;
Blanco, AI ;
Clark, VH .
MEDICAL PHYSICS, 2003, 30 (05) :979-985
[8]
A virtual photon energy fluence model for Monte Carlo dose calculation [J].
Fippel, M ;
Haryanto, F ;
Dohm, O ;
Nüsslin, F ;
Kriesen, S .
MEDICAL PHYSICS, 2003, 30 (03) :301-311
[9]
Development and validation of a BEAMnrc component module for accurate Monte Carlo modelling of the Varian dynamic Millennium multileaf collimator [J].
Heath, E ;
Seuntjens, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2003, 48 (24) :4045-4063
[10]
Large efficiency improvements in BEAMnrc using directional bremsstrahlung splitting [J].
Kawrakow, I ;
Rogers, DWO ;
Walters, BRB .
MEDICAL PHYSICS, 2004, 31 (10) :2883-2898