Performance characterization and first clinical evaluation of a intra-operative compact gamma imager

被引:12
作者
Menard, L
Charon, Y
Solal, M
Ricard, M
Lanièce, P
Mastrippolito, R
Pinot, L
Valentin, L
机构
[1] Inst Phys Nucl, F-91406 Orsay, France
[2] Univ Paris 07, Grp Modelisat Phys Interfaces Biol, F-75005 Paris, France
[3] Inst Gustave Roussy, F-94805 Villejuif, France
[4] INSERM, U66, F-94805 Villejuif, France
关键词
D O I
10.1109/23.819284
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The growing interest of cancer surgeons in intraoperative probes has led to the development of several prototypes of high resolution mini gamma cameras. The aim of this paper is to present a global characterization of the one that we developed and the corresponding first evaluation in a clinical context. The current prototype of POCI (peroperative compact imager) is a 24 mm diameter intensified position sensitive diode optically coupled to a scintillation crystal plate and a novel parallel hole tungsten collimator. In order to face the various clinical situations, two sets of collimator/scintillator imaging heads have been developed either for high spatial resolution or high efficiency purposes. Both of them have first been optimized for Tc-99m labeled tumor detection. Performances of POCI for these two head modules are presented using phantom studies. The results show spatial resolution values ranging between 1 and 1.9 mm (without significant distortion) and a corresponding detection efficiency ranging from 6.10(-3) up to 0.2 events/kBq/sec. Finally, first clinical evaluation of this new high-resolution compact camera concerned sentinel lymph node imaging which is included in melanoma and breast cancer staging protocols. Preliminary results already demonstrate that the performance characteristics of POCI are compatible with intra-operative imaging purposes and suggest how such mini-cameras can improve the success rate of tumor removal surgeries.
引用
收藏
页码:2068 / 2074
页数:7
相关论文
共 19 条
[11]   An original emission tomograph for in vivo brain imaging of small animals [J].
Ochoa, AV ;
Ploux, L ;
Mastrippolito, R ;
Charon, Y ;
Laniece, P ;
Pinot, L ;
Valentin, L .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (04) :1533-1537
[12]   High resolution CsI(Tl)/Si-PIN detector development for breast imaging [J].
Patt, BE ;
Iwanczyk, JS ;
Tull, CR ;
Wang, NW ;
Tornai, MP ;
Hoffman, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (04) :2126-2131
[13]   Development of an intraoperative gamma camera based on a 256-pixel mercuric iodide detector array [J].
Patt, BE ;
Tornai, MP ;
Iwanczyk, JS ;
Levin, CS ;
Hoffman, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1242-1248
[14]  
PERKINS A, 1993, EUR J NUCL MED, V20, P573
[15]   THE IMPACT OF DYNAMIC LYMPHOSCINTIGRAPHY AND GAMMA-PROBE GUIDANCE ON SENTINEL NODE BIOPSY IN MELANOMA [J].
PIJPERS, R ;
COLLET, GJ ;
MEIJER, S ;
HOEKSTRA, OS .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1995, 22 (11) :1238-1241
[16]  
RICARD M, 1993, EUR J NUCL MED, V20, P426
[17]   Evaluation of surgical gamma probes for radioguided sentinel node localisation [J].
Tiourina, T ;
Arends, B ;
Huysmans, D ;
Rutten, H ;
Lemaire, B ;
Muller, S .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1998, 25 (09) :1224-1231
[18]   Discrete scintillator coupled mercuric iodide photodetector arrays for breast imaging [J].
Tornai, MP ;
Patt, BE ;
Iwanczyk, JS ;
Levin, CS ;
Hoffman, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1127-1133
[19]   Investigation of crystal geometries for fiber coupled gamma imaging intra-operative probes [J].
Tornai, MP ;
Levin, CS ;
MacDonald, LR ;
Hoffman, EJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1997, 44 (03) :1254-1261