An efficient calibration method for freehand 3-D ultrasound imaging systems

被引:34
作者
Leotta, DF [1 ]
机构
[1] Univ Washington, Dept Surg, Seattle, WA 98195 USA
关键词
3-D ultrasound; calibration; tracking system; freehand; phantom;
D O I
10.1016/j.ultrasmedbio.2004.05.007
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A phantom has been developed to quickly calibrate a freehand 3-D ultrasound (US) imaging system. Calibration defines the spatial relationship between the US image plane and an external tracking device attached to the scanhead. The phantom consists of a planar array of strings and beads, and a set of out-of-plane strings that guide the user to proper scanhead orientation for imaging. When an US image plane is coincident with the plane defined by the strings, the calibration parameters are calculated by matching of homologous points in the image and phantom. The resulting precision and accuracy of the 3-D imaging system are similar to those achieved with a more complex calibration procedure. The 3-D reconstruction performance of the calibrated system is demonstrated with a magnetic tracking system, but the method could be applied to other tracking devices. (C) 2004 World Federation for Ultrasound in Medicine Biology.
引用
收藏
页码:999 / 1008
页数:10
相关论文
共 10 条
[1]  
AMIN DV, 2001, LECT NOTES COMPUTER, V2208, P940
[2]   LEAST-SQUARES FITTING OF 2 3-D POINT SETS [J].
ARUN, KS ;
HUANG, TS ;
BLOSTEIN, SD .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) :699-700
[3]  
Blackall JM, 2000, LECT NOTES COMPUT SC, V1935, P462
[4]   Calibration of three-dimensional ultrasound images for image-guided radiation therapy [J].
Bouchet, LG ;
Meeks, SL ;
Goodchild, G ;
Bova, FJ ;
Buatti, JM ;
Friedman, WA .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02) :559-577
[5]   3D ULTRASONIC IMAGE FEATURE LOCALIZATION BASED ON MAGNETIC SCANHEAD TRACKING - IN-VITRO CALIBRATION AND VALIDATION [J].
DETMER, PR ;
BASHEIN, G ;
HODGES, T ;
BEACH, KW ;
FILER, EP ;
BURNS, DH ;
STRANDNESS, DE .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1994, 20 (09) :923-936
[6]   Performance of a miniature magnetic position sensor for three-dimensional ultrasound imaging [J].
Leotta, DF ;
Detmer, PR ;
Martin, RW .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1997, 23 (04) :597-609
[7]   Beam calibration without a phantom for creating a 3-D freehand ultrasound system [J].
Muratore, DM ;
Galloway, RL .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (11) :1557-1566
[8]   A fast calibration method for 3-D tracking of ultrasound images using a spatial localizer [J].
Pagoulatos, N ;
Haynor, DR ;
Kim, Y .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2001, 27 (09) :1219-1229
[9]   Rapid calibration for 3-D freehand ultrasound [J].
Prager, RW ;
Rohling, RN ;
Gee, AH ;
Berman, L .
ULTRASOUND IN MEDICINE AND BIOLOGY, 1998, 24 (06) :855-869
[10]   High-definition freehand 3-D ultrasound [J].
Treece, GM ;
Gee, AH ;
Prager, RW ;
Cash, CJC ;
Berman, LH .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2003, 29 (04) :529-546