A Monte Carlo study of radiation transport through multileaf collimators

被引:82
作者
Kim, JO [1 ]
Siebers, JV [1 ]
Keall, PJ [1 ]
Arnfield, MR [1 ]
Mohan, R [1 ]
机构
[1] Virginia Commonwealth Univ, Dept Radiat Oncol, Med Coll Virginia Hosp, Richmond, VA 23298 USA
关键词
Monte Carlo; multileaf collimator;
D O I
10.1118/1.1420734
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Due to the significant increase in the number of monitor units used to deliver a dynamic IMRT treatment, the total MLC leakage (transmission plus scatter) can exceed 10% of the maximum in-field close. To avoid dosimetric errors, this leakage must be accurately accounted for in the dose calculation and conversion of optimized intensity patterns to MLC trajectories used for treatment delivery. In this study, we characterized the leaf end transmission and leakage radiation for Varian 80- and 120-leaf MLCs using Monte Carlo simulations. The complex geometry of the MLC, including the rounded leaf end., leaf edges (tongue-and-groove and offset notch), mounting slots, and holes was modeled using MCNP4b. Studies were undertaken to determine the leakage as a function of field size. components of the leakage, electron contamination, beam hardening and leaf tip effects. The leakage radiation with the MLC configured to fully block the field was determined. Dose for 6 and IS MV beams was calculated at 5 cm depth in a water phantom located at 95 cm SSD, and normalized to the dose for an open field, Dose components were scored separately for radiation transmitted through and scattered from the MLC. For the 80-leaf MLC at 6 MV, the average leakage dose is 1.6%. 1.7%, 1.8%. and 1.9% for 5x5, 10x10, 15x15, and 20x20cm(2) fields, respectively. For the 120-leaf MLC at 6 MV, the average leakage dose is 1.6%, 1.6% 1.7%, and 1.9% for the same field sizes. Measured leakage values for the 120-leaf MLC agreed with calculated values to within 0.1% of the open field dose. The increased leakage with field size is attributed to MLC scattered radiation. The fractional electron contamination for a blocked MLC field is greater than that for an open field. The MLC attenuation significantly affects the photon spectrum, resulting in an increase in percent depth dose at 6 MV, however, little effect is observed at 18 MV. Both phantom scatter and the finite source size contribute to the leaf tip profile observed in phantom. The results of this paper can be applied to fluence-to-trajectory and trajectory-to-fluence calculations for IMRT. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:2497 / 2506
页数:10
相关论文
共 30 条
[1]   A method for determining multileaf collimator transmission and scatter for dynamic intensity modulated radiotherapy [J].
Arnfield, MR ;
Siebers, JV ;
Kim, JO ;
Wu, QW ;
Keall, PJ ;
Mohan, R .
MEDICAL PHYSICS, 2000, 27 (10) :2231-2241
[2]   X-RAY FIELD COMPENSATION WITH MULTILEAF COLLIMATORS [J].
BORTFELD, TR ;
KAHLER, DL ;
WALDRON, TJ ;
BOYER, AL .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1994, 28 (03) :723-730
[3]   Geometric analysis of light-field position of a multileaf collimator with curved ends [J].
Boyer, AL ;
Li, SD .
MEDICAL PHYSICS, 1997, 24 (05) :757-762
[4]  
BRIESMEISTER JF, 1997, LA126257 LOS AL NAT
[5]   THE GENERATION OF INTENSITY-MODULATED FIELDS FOR CONFORMAL RADIOTHERAPY BY DYNAMIC COLLIMATION [J].
CONVERY, DJ ;
ROSENBLOOM, ME .
PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (06) :1359-1374
[6]   Optimization and clinical use of multisegment intensity-modulated radiation therapy for high-dose conformal therapy [J].
Fraass, BA ;
Kessler, ML ;
McShan, DL ;
Marsh, LH ;
Watson, BA ;
Dusseau, WJ ;
Eisbruch, A ;
Sandler, HM ;
Lichter, AS .
SEMINARS IN RADIATION ONCOLOGY, 1999, 9 (01) :60-+
[7]   CHARACTERIZATION OF A MULTILEAF COLLIMATOR SYSTEM [J].
GALVIN, JM ;
SMITH, AR ;
LALLY, B .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (02) :181-192
[8]   SIMULTANEOUS-OPTIMIZATION OF DYNAMIC MULTILEAF COLLIMATION AND SCANNING PATTERNS OR COMPENSATION FILTERS USING A GENERALIZED PENCIL BEAM ALGORITHM [J].
GUSTAFSSON, A ;
LIND, BK ;
SVENSSON, R ;
BRAHME, A .
MEDICAL PHYSICS, 1995, 22 (07) :1141-1156
[9]   DOSIMETRIC CHARACTERISTICS OF A COMMERCIAL MULTILEAF COLLIMATOR [J].
HUQ, MS ;
YU, Y ;
CHEN, ZP ;
SUNTHARALINGAM, N .
MEDICAL PHYSICS, 1995, 22 (02) :241-247
[10]   THE DESIGN AND PERFORMANCE-CHARACTERISTICS OF A MULTILEAF COLLIMATOR [J].
JORDAN, TJ ;
WILLIAMS, PC .
PHYSICS IN MEDICINE AND BIOLOGY, 1994, 39 (02) :231-251