Portal imaging based definition of the planning target volume during pelvic irradiation for gynecological malignancies

被引:15
作者
Mock, U [1 ]
Dieckmann, K [1 ]
Wolff, U [1 ]
Knocke, TH [1 ]
Pötter, R [1 ]
机构
[1] Univ Hosp Vienna, Dept Radiotherapy & Radiolbiol, A-1090 Vienna, Austria
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 1999年 / 45卷 / 01期
关键词
portal imaging; gynecological malignancies; radiation therapy;
D O I
10.1016/S0360-3016(99)00140-6
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Geometrical accuracy in patient positioning can vary substantially during external radiotherapy. This study estimated the set-up accuracy during pelvic irradiation for gynecological malignancies for determination of safety margins (planning target volume, PTV), Methods and Materials: Based on electronic portal imaging devices (EPID), 25 patients undergoing 4-field pelvic irradiation for gynecological malignancies were analyzed with regard to set-up accuracy during the treatment course. Regularly performed EPID images were used in order to systematically assess the systematic and random component of set-up displacements. Anatomical matching of verification and simulation images was followed by measuring corresponding distances between the central axis and anatomical features. Data analysis of set-up errors referred to the x-, y-,and z-axes. Additionally, cumulative frequencies were evaluated, Results: A total of 50 simulation films and 313 verification images were analyzed. For the anterior-posterior (AP) beam direction mean deviations along the x- and z-axes were 1.5 mm and -1.9 mm, respectively, Moreover, random errors of 4.8 mm (x-asis) and 3.0 mm (z-axis) were determined, Concerning the latero-lateral treatment fields, the systematic errors along the two axes were calculated to 2.9 mm (y-axis) and -2.0 mm (z-axis) and random errors of 3.8 mm and 3.5 mm were found, respectively, The cumulative frequency of misalignments less than or equal to 5 mm showed values of 75 % (AP fields) and 72 % (latero-lateral fields). With regard to cumulative frequencies less than or equal to 10 mm quantification revealed values of 97 % for both beam directions. Conclusion: During external pelvic irradiation therapy for gynecological malignancies, EPID images on a regular basis revealed acceptable set-up inaccuracies. Safety margins (PTV) of 1 cm appear to be sufficient, accounting for more than 95% of all deviations. (C) 1999 Elsevier Science Inc.
引用
收藏
页码:227 / 232
页数:6
相关论文
共 25 条
[1]   ONLINE REPOSITIONING DURING TREATMENT OF THE PROSTATE - A STUDY OF POTENTIAL LIMITS AND GAINS [J].
BALTER, JM ;
CHEN, GTY ;
PELIZZARI, CA ;
KRISHNASAMY, S ;
RUBIN, S ;
VIJAYAKUMAR, S .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 27 (01) :137-143
[2]   High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imaging [J].
Bel, A ;
Vos, PH ;
Rodrigus, PTR ;
Creutzberg, CL ;
Visser, AG ;
Stroom, JC ;
Lebesque, JV .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (02) :321-332
[3]   A VERIFICATION PROCEDURE TO IMPROVE PATIENT SET-UP ACCURACY USING PORTAL IMAGES [J].
BEL, A ;
VANHERK, M ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :253-260
[4]   MAXIMIZING SETUP ACCURACY USING PORTAL IMAGES AS APPLIED TO A CONFORMAL BOOST TECHNIQUE FOR PROSTATIC-CANCER [J].
BIJHOLD, J ;
LEBESQUE, JV ;
HART, AAM ;
VIJLBRIEF, RE .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (04) :261-271
[5]  
BYHARDT R, 1987, INT J RADIAT ONCOL, V4, P881
[6]   ROLE OF PORTAL IMAGING IN CLINICAL RADIOTHERAPY - FLORENCE EXPERIENCE [J].
CIONINI, L ;
BUCCIOLINI, M .
RADIOTHERAPY AND ONCOLOGY, 1993, 29 (02) :230-236
[7]   ROUTINE CLINICAL ONLINE PORTAL IMAGING FOLLOWED BY IMMEDIATE FIELD ADJUSTMENT USING A TELE-CONTROLLED PATIENT COUCH [J].
DENEVE, W ;
VANDENHEUVEL, F ;
DEBEUKELEER, M ;
COGHE, M ;
THON, L ;
DEROOVER, P ;
VANLANCKER, M ;
STORME, G .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (01) :45-54
[8]   INTERACTIVE USE OF ONLINE PORTAL IMAGING IN PELVIC RADIATION [J].
DENEVE, W ;
VANDENHEUVEL, F ;
COGHE, M ;
VERELLEN, D ;
DEBEUKELEER, M ;
ROELSTRAETE, A ;
DEROOVER, P ;
THON, L ;
STORME, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03) :517-524
[9]   TIME TREND OF PATIENT SETUP DEVIATIONS DURING PELVIC IRRADIATION USING ELECTRONIC PORTAL IMAGING [J].
ELGAYED, AAH ;
BEL, A ;
VIJLBRIEF, R ;
BARTELINK, H ;
LEBESQUE, JV .
RADIOTHERAPY AND ONCOLOGY, 1993, 26 (02) :162-171
[10]   REPRODUCIBILITY OF PATIENT POSITIONING DURING ROUTINE RADIOTHERAPY, AS ASSESSED BY AN INTEGRATED MEGAVOLTAGE IMAGING-SYSTEM [J].
GILDERSLEVE, J ;
DEARNALEY, DP ;
EVANS, PM ;
SWINDELL, W .
RADIOTHERAPY AND ONCOLOGY, 1995, 35 (02) :151-160