Quality assurance of a system for improved target localization and patient set-up that combines real-time infrared tracking and stereoscopic X-ray imaging

被引:128
作者
Verellen, D [1 ]
Soete, G [1 ]
Linthout, N [1 ]
Van Acker, S [1 ]
De Roover, P [1 ]
Vinh-Hung, V [1 ]
Van de Steene, J [1 ]
Storme, G [1 ]
机构
[1] Free Univ Brussels, Acad Hosp, Dept Radiotherapy, Ctr Oncol, B-1090 Brussels, Belgium
关键词
automated patient positioning; intra-fractional correction of three-dimensional set-up errors; infrared markers; stereoscopic X-ray imaging;
D O I
10.1016/S0167-8140(02)00385-7
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: The aim of this study is to investigate the positional accuracy of a prototype X-ray imaging tool in combination with a real-time infrared tracking device allowing automated patient set-up in three dimensions. Material and methods: A prototype X-ray imaging tool has been integrated with a commercially released real-time infrared tracking device. The system, consisting of two X-ray tubes mounted to the ceiling and a centrally located amorphous silicon detector has been developed for automated patient positioning from outside the treatment room prior to treatment. Two major functions are supported: (a) automated fusion of the actual treatment images with digitally reconstructed radiographs (DRRs) representing the desired position; (b) matching of implanted radio opaque markers. Measurements of known translational (up to 30.0 mm) and rotational (up to 4.0degrees) set-up errors in three dimensions as well as hidden target tests have been performed on anthropomorphic phantoms. Results: The system's accuracy can be represented with the mean three-dimensional displacement vector, which yielded 0.6 mm (with an overall SD of 0.9 mm) for the fusion of DRRs and X-ray images. Average deviations between known translational errors and calculations varied from - 0.3 to 0.6 mm with a standard deviation in the range of 0.6-1.2 mm. The marker matching algorithm yielded a three-dimensional uncertainty of 0.3 rum (overall SD: 0.4 mm), with averages ranging from 0.0 to 0.3 mm and a standard deviation in the range between 0.3 and 0.4 turn. Conclusions: The stereoscopic X-ray imaging device integrated with the real-time infrared tracking device represents a positioning tool allowing for the geometrical accuracy that is required for conformal radiation therapy of abdominal and pelvic lesions, within an acceptable time-frame. (C) 2002 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:129 / 141
页数:13
相关论文
共 47 条
[1]   Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate [J].
Alasti, H ;
Petric, MP ;
Catton, CN ;
Warde, PR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03) :869-884
[2]   Automated localization of the prostate at the time of treatment using implanted radiopaque markers: Technical feasibility [J].
Balter, JM ;
Lam, KL ;
Sandler, HM ;
Littles, JF ;
Bree, RL ;
TenHaken, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (05) :1281-1286
[3]   MEASUREMENT OF PROSTATE MOVEMENT OVER THE COURSE OF ROUTINE RADIOTHERAPY USING IMPLANTED MARKERS [J].
BALTER, JM ;
SANDLER, HM ;
LAM, K ;
BREE, RL ;
LICHTER, AS ;
TENHAKEN, RK .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 31 (01) :113-118
[4]   Target margins for random geometrical treatment uncertainties in conformal radiotherapy [J].
Bel, A ;
vanHerk, M ;
Lebesque, JV .
MEDICAL PHYSICS, 1996, 23 (09) :1537-1545
[5]   A computerized remote table control for fast on-line patient repositioning: Implementation and clinical feasibility [J].
Bel, A ;
Petrascu, O ;
Van de Vondel, I ;
Coppens, L ;
Linthout, N ;
Verellen, D ;
Storme, G .
MEDICAL PHYSICS, 2000, 27 (02) :354-358
[6]   High-precision prostate cancer irradiation by clinical application of an offline patient setup verification procedure, using portal imaging [J].
Bel, A ;
Vos, PH ;
Rodrigus, PTR ;
Creutzberg, CL ;
Visser, AG ;
Stroom, JC ;
Lebesque, JV .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1996, 35 (02) :321-332
[7]   A DIAGNOSTIC X-RAY FIELD VERIFICATION DEVICE FOR A 10 MV LINEAR-ACCELERATOR [J].
BIGGS, PJ ;
GOITEIN, M ;
RUSSELL, MD .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1985, 11 (03) :635-643
[8]   A REVIEW OF ELECTRONIC PORTAL IMAGING DEVICES (EPIDS) [J].
BOYER, AL ;
ANTONUK, L ;
FENSTER, A ;
VANHERK, M ;
MEERTENS, H ;
MUNRO, P ;
REINSTEIN, LE ;
WONG, J .
MEDICAL PHYSICS, 1992, 19 (01) :1-16
[9]   ROUTINE CLINICAL ONLINE PORTAL IMAGING FOLLOWED BY IMMEDIATE FIELD ADJUSTMENT USING A TELE-CONTROLLED PATIENT COUCH [J].
DENEVE, W ;
VANDENHEUVEL, F ;
DEBEUKELEER, M ;
COGHE, M ;
THON, L ;
DEROOVER, P ;
VANLANCKER, M ;
STORME, G .
RADIOTHERAPY AND ONCOLOGY, 1992, 24 (01) :45-54
[10]   INTERACTIVE USE OF ONLINE PORTAL IMAGING IN PELVIC RADIATION [J].
DENEVE, W ;
VANDENHEUVEL, F ;
COGHE, M ;
VERELLEN, D ;
DEBEUKELEER, M ;
ROELSTRAETE, A ;
DEROOVER, P ;
THON, L ;
STORME, G .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1993, 25 (03) :517-524