In vivo calibration of the subcutaneous amperometric glucose sensors using a non-enzyme electrode

被引:19
作者
Jeong, RA
Hwang, JY
Joo, S
Chung, TD
Park, S
Kang, SK
Lee, WY
Kim, HC
机构
[1] Seoul Natl Univ, Coll Med, Dept Biomed Engn, Med Res Ctr, Seoul 110744, South Korea
[2] Seoul Natl Univ, Inst Med & Biol Engn, Med Res Ctr, Seoul 110744, South Korea
[3] Sungshin Womens Univ, Dept Chem, Seoul 136742, South Korea
[4] Yonsei Univ, Dept Chem, Seoul 120749, South Korea
关键词
implantable continuous glucose sensor; two-point calibration; perfluorinated polymer; telemetry;
D O I
10.1016/S0956-5663(03)00219-7
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A new two-point calibration method for the subcutaneous amperometric continuous glucose sensor is reported. The proposed method is based on direct measurement of the background current (I-o) using a non-enzyme electrode. For in vivo test, three electrodes were implanted in rabbits. Two of the three were identical needle-type enzyme electrodes with perfluorinated polymer outer layers (Pt/enzyme layer/Kel-F/PTFE/Kel-F/Nafion) that were placed in subcutaneous tissue and in a vessel (ear artery), respectively. And one non-enzyme electrode with exactly the same membrane composition as those of other two was in the subcutaneous layer to measure the background current. Implantation in the subcutaneous layer generated many crevices on the protecting layers of the electrodes. The signals from enzyme electrodes were effectively corrected by the measured background current from the non-enzyme electrode. In addition, a telemetric monitoring system was developed and evaluated for in vivo continuous glucose monitoring in order to alleviate the problems of motion artifact. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:313 / 319
页数:7
相关论文
共 15 条
  • [1] Interstitial glucose concentration and glycemia: implications for continuous subcutaneous glucose monitoring
    Aussedat, B
    Dupire-Angel, M
    Gifford, R
    Klein, JC
    Wilson, GS
    Reach, G
    [J]. AMERICAN JOURNAL OF PHYSIOLOGY-ENDOCRINOLOGY AND METABOLISM, 2000, 278 (04): : E716 - E728
  • [2] Calibration of a subcutaneous amperometric glucose sensor implanted for 7 days in diabetic patients - Part 2. Superiority of the one-point calibration method
    Choleau, C
    Klein, JC
    Reach, G
    Aussedat, B
    Demaria-Pesce, V
    Wilson, GS
    Gifford, R
    Ward, WK
    [J]. BIOSENSORS & BIOELECTRONICS, 2002, 17 (08) : 647 - 654
  • [3] Calibration of a subcutaneous amperometric glucose sensor - Part 1. Effect of measurement uncertainties on the determination of sensor sensitivity and background current
    Choleau, C
    Klein, JC
    Reach, G
    Aussedat, B
    Demaria-Pesce, V
    Wilson, GS
    Gifford, R
    Ward, WK
    [J]. BIOSENSORS & BIOELECTRONICS, 2002, 17 (08) : 641 - 646
  • [4] Reproducible fabrication of miniaturized glucose sensors: preparation of sensing membranes for continuous monitoring
    Chung, TD
    Jeong, RA
    Kang, SK
    Kim, HC
    [J]. BIOSENSORS & BIOELECTRONICS, 2001, 16 (9-12) : 1079 - 1087
  • [5] EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE
    CLARKE, WL
    COX, D
    GONDERFREDERICK, LA
    CARTER, W
    POHL, SL
    [J]. DIABETES CARE, 1987, 10 (05) : 622 - 628
  • [6] Reservations on the use of error grid analysis for the validation of blood glucose assays
    Gough, DA
    Botvinick, EL
    [J]. DIABETES CARE, 1997, 20 (06) : 1034 - 1036
  • [7] TOWARDS CONTINUOUS GLUCOSE MONITORING - INVIVO EVALUATION OF A MINIATURIZED GLUCOSE SENSOR IMPLANTED FOR SEVERAL DAYS IN RAT SUBCUTANEOUS TISSUE
    MOATTISIRAT, D
    CAPRON, E
    POITOUT, V
    REACH, G
    BINDRA, DS
    ZHANG, Y
    WILSON, GS
    THEVENOT, DR
    [J]. DIABETOLOGIA, 1992, 35 (03) : 224 - 230
  • [8] PERFORMANCE OF SUBCUTANEOUSLY IMPLANTED NEEDLE-TYPE GLUCOSE SENSORS EMPLOYING A NOVEL TRILAYER COATING
    MOUSSY, F
    HARRISON, DJ
    OBRIEN, DW
    RAJOTTE, RV
    [J]. ANALYTICAL CHEMISTRY, 1993, 65 (15) : 2072 - 2077
  • [9] Continuous glucose monitoring in the free-moving rat
    Thomé-Duret, V
    Aussedat, B
    Reach, G
    Gangnerau, MN
    Lemonnier, F
    Klein, JC
    Zhang, Y
    Hu, Y
    Wilson, GS
    [J]. METABOLISM-CLINICAL AND EXPERIMENTAL, 1998, 47 (07): : 799 - 803
  • [10] Use of a subcutaneous glucose sensor to detect decreases in glucose concentration prior to observation in blood
    ThomeDuret, V
    Reach, G
    Gangnerau, MN
    Lemonnier, F
    Klein, JC
    Zhang, YN
    Hu, YB
    Wilson, GS
    [J]. ANALYTICAL CHEMISTRY, 1996, 68 (21) : 3822 - 3826