Achieving the full performance of highly efficient columns by optimizing conventional benchmark high-performance liquid chromatography instruments

被引:156
作者
Gritti, Fabrice [1 ]
Sanchez, Carl A. [2 ]
Farkas, Tivadar [2 ]
Guiochon, Georges [1 ]
机构
[1] Univ Tennessee, Dept Chem, Knoxville, TN 37996 USA
[2] Phenomenex Inc, Torrance, CA 90501 USA
基金
美国国家科学基金会;
关键词
Extra-column effects; Variance; Instrumentation; Column efficiency; Resolutions; Agilent; 1100; 1200; Agilent 1290 Infinity HPLC; Acquity; Resolution; Column packing technology; Shell particles; Kinetex-C(18); DISPERSION; PARTICLES; TEMPERATURE; PRESSURE; VOLUME; GAS;
D O I
10.1016/j.chroma.2010.02.044
中图分类号
Q5 [生物化学];
学科分类号
070307 [化学生物学];
摘要
A series of experiments and measurements demonstrate the importance of minimizing the extra-column band broadening contribution of the instrument used. The combination of several measures allowed the achievement of the full potential efficiency of three Kinetex-C(18) columns, using a conventional liquid chromatograph. The first measure consists in minimizing the extra-column volume of the instrument, without increasing much its back pressure contribution, by changing the needle seat volume, the inner diameter and length of the capillary connectors, and the volume of the detector cell of a standard instrument (Agilent 1100). The second measure consists in injecting a volume of weak eluent (less than half the elution strength of the mobile phase) right after the sample, before the sample had time to reach the column. Experimental results show that these changes could provide most of the resolution expected from the true column performance. After the changes were made, the resolutions of the 2.1 mm x 50 mm, 4.6 mm x 50 mm, and 4.6 mm x 100 mm Kinetex-C(18) columns for compounds having retention factors close to 1 were increased by about 180, 35, and 30%, respectively. The resolutions obtained are then similar to those measured with advanced instruments like the Agilent 1200, the Agilent 1290 Infinity HPLC, and the Acquity chromatographs. (C) 2010 Elsevier B.V. All rights reserved.
引用
收藏
页码:3000 / 3012
页数:13
相关论文
共 18 条
[1]
[Anonymous], 2008, 32 INT S HIGH PERF L
[2]
DISPERSION OF PEAKS BY SHORT STRAIGHT OPEN TUBES IN LIQUID-CHROMATOGRAPHY SYSTEMS [J].
ATWOOD, JG ;
GOLAY, MJE .
JOURNAL OF CHROMATOGRAPHY, 1981, 218 (1-3) :97-122
[3]
EXTRA-COLUMN EFFECTS IN HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY .1. THEORETICAL-STUDY OF THE INJECTION PROBLEM [J].
COLIN, H ;
MARTIN, M ;
GUIOCHON, G .
JOURNAL OF CHROMATOGRAPHY, 1979, 185 (DEC) :79-95
[4]
Characteristics of superficially-porous silica particles for fast HPLC:: Some performance comparisons with sub-2-μm particles [J].
DeStefano, J. J. ;
Langlois, T. J. ;
Kirkland, J. J. .
JOURNAL OF CHROMATOGRAPHIC SCIENCE, 2008, 46 (03) :254-260
[5]
Effects of extra-column band spreading, liquid chromatography system operating pressure, and column temperature on the performance of sub-2-μm porous particles [J].
Fountain, Kenneth J. ;
Neue, Uwe D. ;
Grumbach, Eric S. ;
Diehl, Diane M. .
JOURNAL OF CHROMATOGRAPHY A, 2009, 1216 (32) :5979-5988
[6]
EARLY PHASES OF THE DISPERSION OF A SAMPLE INJECTED IN POISEUILLE FLOW [J].
GOLAY, MJE ;
ATWOOD, JG .
JOURNAL OF CHROMATOGRAPHY, 1979, 186 (DEC) :353-370
[7]
The ultimate band compression factor in gradient elution chromatography [J].
Gritti, Fabrice ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2008, 1178 (1-2) :79-91
[8]
Influence of the errors made in the measurement of the extra-column volume on the accuracies of estimates of the column efficiency and the mass transfer kinetics parameters [J].
Gritti, Fabrice ;
Felinger, Attila ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2006, 1136 (01) :57-72
[9]
Performance of columns packed with the new shell particles, Kinetex-C18 [J].
Gritti, Fabrice ;
Leonardis, Irene ;
Shock, David ;
Stevenson, Paul ;
Shalliker, Andrew ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (10) :1589-1603
[10]
Performance of new prototype packed columns for very high pressure liquid chromatography [J].
Gritti, Fabrice ;
Guiochon, Georges .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (09) :1485-1495