Potential and limitations of the automatic detection of fiducial markers using an amorphous silicon flat-panel imager

被引:18
作者
Buck, D [1 ]
Alber, M [1 ]
Nüsslin, F [1 ]
机构
[1] Univ Tubingen, Klin Radioonkol, Abt Med Phys, D-72076 Tubingen, Germany
关键词
D O I
10.1088/0031-9155/48/6/305
中图分类号
R318 [生物医学工程];
学科分类号
0831 [生物医学工程];
摘要
Amorphous silicon electronic portal imaging devices (a-Si EPIDs) allow fast acquisition of high resolution portal images (PI). A visualization of organ movement for adaptive image-guided radiotherapy (IGRT) can be reached by implantation and automatic detection of fiducial markers. A method of automatic detection has been developed for fiducial spherical tungsten markers on PIs, acquired with an a-Si flat-panel imager. The detection method consists of a 2D Mexican hat filter (MHF), whose parameters are tuned to the particular marker signal. The high selectivity of this filter allows a reliable and precise detection of tungsten markers down to a diameter of 1.5 mm. The presented method allows fast, automatic and unsupervised detection of markers. Inevitably, the detection is hampered by image background (bone structures, etc) and noise. A detection success rate higher than 95% was reached, analysing PIs of patients with markers fixed on their skin. Furthermore, this approach to automatic marker detection can also be generalized to elliptic MHFs for the detection of cylindrical markers. The accuracy and detection probability of this method may allow accurate and fast online localization of the considered organ.
引用
收藏
页码:763 / 774
页数:12
相关论文
共 12 条
[1]
Portal imaging for evaluation of daily on-line setup errors and off-line organ motion during conformal irradiation of carcinoma of the prostate [J].
Alasti, H ;
Petric, MP ;
Catton, CN ;
Warde, PR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2001, 49 (03) :869-884
[2]
OPTIMIZATION OF AUTOMATIC PORTAL IMAGE-ANALYSIS [J].
GILHUIJS, KGA ;
TOUW, A ;
VANHERK, M ;
VIJLBRIEF, RE .
MEDICAL PHYSICS, 1995, 22 (07) :1089-1099
[3]
Haase M, 2000, PARADIGMS OF COMPLEXITY, P287
[4]
Fiducial-based targeting accuracy for external-beam radiotherapy [J].
Murphy, MJ .
MEDICAL PHYSICS, 2002, 29 (03) :334-344
[5]
Detection of fiducial gold markers for automatic on-line megavoltage position verification using a marker extraction kernel (MEK) [J].
Nederveen, A ;
Lagendijk, J ;
Hofman, P .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2000, 47 (05) :1435-1442
[6]
Measurements and clinical consequences of prostate motion during a radiotherapy fraction [J].
Nederveen, AJ ;
van der Heide, UA ;
Dehnad, H ;
van Moorselaar, RJA ;
Hofman, P ;
Lagendijk, JJW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2002, 53 (01) :206-214
[7]
Feasibility of automatic marker detection with an a-Si flat-panel imager [J].
Nederveen, AJ ;
Lagendijk, JJW ;
Hofman, P .
PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (04) :1219-1230
[8]
RIEDER A., 1998, WAVELETS THEORIE ANW
[9]
Automated seed detection and three-dimensional reconstruction. II. Reconstruction of permanent prostate implants using simulated annealing [J].
Tubic, D ;
Zaccarin, A ;
Beaulieu, L ;
Pouliot, J .
MEDICAL PHYSICS, 2001, 28 (11) :2272-2279
[10]
Automated seed detection and three-dimensional reconstruction. I. Seed localization from fluoroscopic images or radiographs [J].
Tubic, D ;
Zaccarin, A ;
Pouliot, J ;
Beaulieu, L .
MEDICAL PHYSICS, 2001, 28 (11) :2265-2271